TABLE 2.

Precision testing of PolyTiter unit values

Specimen type and runAvg of three PolyTiter reads/wellSingle PolyTiter read/well
Reader 1 (avg)Reader 2 (avg)Overall (avg)Source of variationCV (%)Reader 1 (avg)Reader 2 (avg)Overall (avg)Source of variationCV (%)
Low40.8440.41
    Run 136.1743.27Between-reader5.4835.5041.60Between-reader5.70
    Run 239.8343.53Between-run3.9839.5043.40Between-run3.59
    Run 339.8342.47Within-run3.8040.5042.20Within-run6.24
    Run 441.5040.2040.0040.40
    Run 538.5043.0738.0043.00
    Avg39.1742.51Total7.7638.7042.12Total9.18
Medium55.7755.62
    Run 149.0058.00Between-reader6.3349.0058.60Between-reader6.88
    Run 256.5056.73Between-run4.7556.0056.60Between-run4.30
    Run 355.8358.60Within-run3.3155.5059.00Within-run4.46
    Run 454.8360.4054.0060.00
    Run 549.5058.3349.5058.00
    Avg53.1358.41Total8.5852.8058.44Total9.26
High64.4164.14
    Run 166.3365.87Between-reader5.1466.5065.20Between-reader5.39
    Run 261.1766.93Between-run2.6960.5066.80Between-run3.21
    Run 361.3367.07Within-run1.9962.0067.00Within-run3.28
    Run 460.3366.8759.0067.20
    Run 560.8367.3360.0067.20
    Avg62.0066.81Total6.1361.6066.68Total7.08