Table 1.

Intra-assay precision of the 1A7-IL-2 ELISAa

MeasurementResult for sample
123
No. of determinations202020
Mean concn (ng/ml)14.462.991.8
CV (%)5.66.36.2
  • a Three samples of known concentrations of the fusion protein were assayed 20 times on one plate. The obtained mean concentration was calculated as well as the coefficient of variation (defined as the standard deviation/mean).