Skip to main content
  • ASM
    • Antimicrobial Agents and Chemotherapy
    • Applied and Environmental Microbiology
    • Clinical Microbiology Reviews
    • Clinical and Vaccine Immunology
    • EcoSal Plus
    • Eukaryotic Cell
    • Infection and Immunity
    • Journal of Bacteriology
    • Journal of Clinical Microbiology
    • Journal of Microbiology & Biology Education
    • Journal of Virology
    • mBio
    • Microbiology and Molecular Biology Reviews
    • Microbiology Resource Announcements
    • Microbiology Spectrum
    • Molecular and Cellular Biology
    • mSphere
    • mSystems
  • Log in
  • My Cart

Main menu

  • Home
  • Articles
    • Archive
  • About the Journal
    • About CVI
    • For Librarians
    • For Advertisers
    • FAQ
  • ASM
    • Antimicrobial Agents and Chemotherapy
    • Applied and Environmental Microbiology
    • Clinical Microbiology Reviews
    • Clinical and Vaccine Immunology
    • EcoSal Plus
    • Eukaryotic Cell
    • Infection and Immunity
    • Journal of Bacteriology
    • Journal of Clinical Microbiology
    • Journal of Microbiology & Biology Education
    • Journal of Virology
    • mBio
    • Microbiology and Molecular Biology Reviews
    • Microbiology Resource Announcements
    • Microbiology Spectrum
    • Molecular and Cellular Biology
    • mSphere
    • mSystems

User menu

  • Log in
  • My Cart

Search

  • Advanced search
Clinical and Vaccine Immunology
publisher-logosite-logo

Advanced Search

  • Home
  • Articles
    • Archive
  • About the Journal
    • About CVI
    • For Librarians
    • For Advertisers
    • FAQ
CLINICAL LABORATORY IMMUNOLOGY

Measuring Immunoglobulin G Antibodies to Tetanus Toxin, Diphtheria Toxin, and Pertussis Toxin with Single-Antigen Enzyme-Linked Immunosorbent Assays and a Bead-Based Multiplex Assay

Sabine Reder, Marion Riffelmann, Christian Becker, Carl Heinz Wirsing von König
Sabine Reder
1Institut Virion\Serion GmbH, Würzburg, Germany
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Marion Riffelmann
2Institut für Infektiologie Krefeld GmbH, Krefeld, Germany
3HELIOS Klinikum Krefeld, Krefeld, Germany
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Christian Becker
3HELIOS Klinikum Krefeld, Krefeld, Germany
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Carl Heinz Wirsing von König
2Institut für Infektiologie Krefeld GmbH, Krefeld, Germany
3HELIOS Klinikum Krefeld, Krefeld, Germany
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • For correspondence: carlheinz.wirsingvonkoenig@helios-kliniken.de
DOI: 10.1128/CVI.00225-07
  • Article
  • Figures & Data
  • Info & Metrics
  • PDF
Loading

ABSTRACT

Bead-based assay systems offer the possibility of measuring several specific antibodies in one sample simultaneously. This study evaluated a vaccine panel of a multianalyte system that measures antibodies to tetanus toxin, diphtheria toxin, and pertussis toxin (PT) from Bordetella pertussis. The antibody concentrations of human immunoglobulin G (IgG) to PT, tetanus toxin, and diphtheria toxin were measured in 123 serum pairs (total of 246 sera) from a vaccine study. The multianalyte bead assay was compared to a standardized in-house IgG- anti-PT enzyme-linked immunosorbent assay (ELISA) of the German reference laboratory for bordetellae, as well as to various commercially available ELISAs for anti-PT IgG, anti-tetanus IgG, and anti-diphtheria IgG. The results of the multiplex assay regarding the antibodies against diphtheria toxin compared favorably with a regression coefficient of 0.938 for values obtained with an ELISA from the same manufacturer used as a reference. Similarly, antibodies to tetanus toxin showed a correlation of 0.910 between the reference ELISA and the multianalyte assay. A correlation coefficient of 0.905 was found when an “in-house” IgG anti-PT and the multiplex assay were compared. Compared to single ELISA systems from two other manufacturers, the multiplex assay performed similarly well or better. The multianalyte assay system was a robust system with fast and accurate results, analyzing three parameters simultaneously in one sample. The system was well suited to quantitatively determine relevant vaccine induced antibodies compared to in-house and commercially available single-antigen ELISA systems.

Measuring antibodies to vaccine antigens after immunization is necessary for some indications. Vaccines for primary immunization of infants, as well as vaccines for use in older children, adolescents, and adults, are now mostly combination products. The primary combination always consists of a tetanus, diphtheria and accellular pertussis component. The pertussis component may contain various antigens, such as pertussis toxin (PT), filamentous hemagglutinin (FHA), pertactin, and/or fimbriae. PT, however, is thought to be the most important single component and is included in all licensed vaccines.

Immunoglobulin G (IgG) antibodies to tetanus toxin, diphtheria toxin, and PT in human serum are mostly measured by enzyme-linked immunosorbent assay (ELISA) (2, 8, 9), requiring the coating of a single vaccine antigen on a solid phase. These tests are somewhat labor-intensive and time-consuming and require substantial amounts of purified antigens. With the introduction of multiplexed assay systems using different beads as a solid phase for flow cytometry, it is now possible to measure antibodies to various antigens in a single assay (5, 6, 7, 10).

In this study, we validated a multiplexed bead assay against three conventional ELISA systems for every antigen.

MATERIALS AND METHODS

Serum samples.Anonymous surplus samples from a vaccine study were used. In the present study, 9- to 16-year-old adolescents with a low titer of IgG anti-PT, no history of pertussis immunization, and no history of pertussis received a single dose of an acellular combination vaccine that is licensed for adolescents and adults in Germany (Boostrix; GlaxoSmithKline Pharma GmbH & Co. KG, Munich, Germany) (3). Paired serum samples were taken before and 29 to 49 days after vaccination.

Methods.IgG antibodies to tetanus toxin were measured by three commercially available ELISAs and a multiplex bead assay: Serion ELISA classic tetanus (Virion\Serion, Würzburg, Germany), Virotech ELISA tetanus (Virotech, Rüsselsheim, Germany), and anti-tetanus toxoid ELISA Euroimmun (Euroimmun, Lübeck, Germany), and Serion Multianalyt diphtheria/tetanus/Bordetella pertussis toxin (Virion\Serion).

IgG antibodies against diphtheria toxin were measured by three commercially available ELISAs and the multiplex bead assay: Serion ELISA classic Diphtherie (Virion\Serion), Virotech ELISA Diphtherie (Virotech), Anti-Diphtherie-Toxoid ELISA Euroimmun (Euroimmun), and Serion Multianalyt diphtheria/tetanus/B. pertussis toxin (Virion\Serion).

IgG antibodies to pertussis toxin were measured by an in-house ELISA, two commercially available ELISAs, and the multiplex bead assay: in-house ELISA systems separately measuring IgG-antibodies against PT, FHA, and pertactin (11); Serion ELISA classic B. pertussis toxin (Virion\Serion), Virotech ELISA pertussis toxin IgG (Virotech), and Serion Multianalyt diphtheria/tetanus/Bordetella pertussis toxin (Virion\Serion).

Reference methods.The following ELISA systems were defined as reference methods for the present study: IgG anti-tetanus toxin, Serion ELISA classic; IgG anti-diphtheria toxin, Serion ELISA classic; and IgG anti-PT, standardized in-house ELISA system (11). ELISAs for measuring antibodies against tetanus toxoid and diphtheria toxoid were previously validated against the Vero cell assay as the reference method (9), and for tetanus toxoid an in vivo mouse test was used as a reference method (8). No reference method exists for measuring antibodies to PT.

All ELISAs were run on a semiautomated washer, read on an E-max microplate reader (Molecular Devices, Ismaning, Germany), and evaluated according to the manufacturer's instructions. The in-house assay was evaluated by using a four-parameter logistic regression. Samples were diluted according to the manufacturers' instructions. Samples with a value above the range of the standard curve were further diluted according to the manufacturers' instructions.

Multiplex bead assay.The multiplex bead assay is based on a proprietary method patented in Germany in 1983 (7), and it has been used for the detection of antibodies to tetanus toxoid, diphtheria toxoid, and B. pertussis antigens (5, 6). In brief, the method is based on microparticles of different sizes (e.g., 4.0 and 5.5 μm in diameter), which are labeled with a red fluorescent dye at different intensities. Target antigens, i.e., purified tetanus toxoid, purified diphtheria toxoid, and purified pertussis toxin, are coupled to the microparticles by covalent peptide bonding. The test also contains two different control particles that have a different fluorescent dye and serve as a function control for the test and for the cytometer. The test is done in microtiter filter plates. A serum dilution is mixed with the microparticles and incubated for 30 min. The cavities are emptied by vacuum and washed four times. A second antibody to human IgG coupled to phycoerythrin is added, followed by incubation for another 30 min. The cavities are emptied by vacuum again and washed four times. Particles are resuspended by gentle mixing (5 min for 600 rpm on a horizontal shaker). The beads from every cavity are then measured with a flow cytometer. In our study, the test was run on an EPICS XL-MCL cytometer (Beckman-Coulter, Krefeld, Germany), and it was evaluated with assay-specific software (Serion Multianalyt evaluation software). The test format has also been validated for use with other flow cytometers, such as CyFlow (Partec Corp.) and FACSCalibur (BD Biosciences).

Reference preparations.The ELISAs used were based on the following reference preparations according to the information of the manufacturer.

IgG anti-tetanus toxin.All ELISAs and the multiplex bead assay refer to the World Health Organization (WHO) reference preparation, which is referred to differently by the various manufacturers (Multianalyt system, NIBSC code 76/589; Serion ELISA, 1.Intern.Standard, code TE-3; Virotech ELISA, WHO TE-3; Euroimmun ELISA, NIBSC code 76/589).

IgG anti-diphtheria toxin.The ELISAs and the multiplex bead assay refer to reference preparations by either WHO or by the Statens Seruminstitut, which are referred to differently by the manufacturers (Multianalyt system, NIBSC code 00/496; Serion ELISA, 1.Intern.Standard, Statens Seruminstitut, Copenhagen, Denmark; Virotech ELISA, NIBSC code 00/496; Euroimmun ELISA, NIBSC code 91/534.).

IgG anti-PT in-house ELISA, multianalyte system, and Serion ELISA.Values are measured in ELISA units (EU)/ml and refer to reference preparation 3 from the Laboratory of Pertussis, CBER, Food and Drug Administration (FDA), Bethesda, MD. The Virotech ELISA reports results as arbitrary units/ml (VE/ml), which should correlate to the CBER/FDA reference preparation.

Lower levels of detection.Lower levels of detection were as follows: (i) for IgG anti-tetanus toxin, Serion ELISA (0.1 IU/ml), Serion Multianalyt (0.06 IU/ml), Virotech ELISA (0.1 IU/ml), and Euroimmun ELISA (0.01 IU/ml); (ii) for IgG anti-diphtheria toxin, Serion ELISA (0.05 IU/ml), Serion Multianalyt (0.025 IU/ml), Virotech ELISA (0.1 IU/ml), and Euroimmun ELISA (0.01 IU/ml); and (iii) for IgG anti-PT, in-house ELISA (2 EU/ml), Virion-Serion classic ELISA (5 EU/ml), Serion Multianalyt (4 EU/ml), and Virotech ELISA (1 VE/ml).

Range of primary standard curves.The ranges of primary standard curves were as follows: (i) for IgG anti-tetanus toxin, Serion ELISA (0.1 to 5.0 IU/ml), Serion Multianalyt (0.06 to 6.0 IU/ml), Virotech ELISA (0.1 to 5.0 IU/ml), and Euroimmun ELISA (0.01 to 10.0 IU/ml); (ii) for IgG anti-diphtheria toxin, Serion ELISA (0.05 to 2.0 IU/ml), Serion Multianalyt (0.025 to 6.0 IU/ml), Virotech ELISA (0.10 to 5.0 IU/ml), and Euroimmun ELISA (0.01 to 2.0 IU/ml); and (iii) for IgG anti-PT, in-house ELISA (2 to 200 EU/ml), Serion ELISA (5 to 600 EU/ml), Serion Multianalyt (4 to 500 EU/ml), and Virotech ELISA (not defined [0.6 to 51.5 VE/ml were used]).

Statistical evaluation. (i) Regression analysis.Scatter plots were constructed from primary data using the following adjustments. Values below the lower limit of detection were given 50% of the numeric value of the lower limit of detection. Values above the range of the standard curve (including secondary dilutions) were given the highest measurable value.

Scatter plots were made by using Sigma-Plot (SPSS, Jandel Scientific, Erkrath, Germany) including a linear regression line. In addition, the 95% prediction intervals were also plotted. For some comparisons, scatter plots from all results were constructed by using the quantitative values of all ELISAs and the multianalyte system with different ranges. These ranges were as follows: for IgG anti-tetanus toxin <1.0 IU/ml, 1.0 to 2.0 IU/ml, 2.0 to 10.0 IU/ml, and >10.0 IU/ml; for IgG anti-diphtheria toxin, <0.1 IU/ml, 0.1 to 1.0 IU/ml, 1.0 to 2.0 IU/ml, and >2.0 IU/ml; and for IgG anti-PT, <50 EU/ml, 50 to 100 EU/ml, and >100 EU/ml. Bland-Altman plots were constructed by using Prism software (GraphPad Software, San Diego, CA).

(ii) Relative concordance of ranges.In order to determine the relative concordance of ranges, it is important to compare ranges of antibodies, since these may be clinically relevant to the interpretation of results. The following ranges were defined and compared. For IgG anti-tetanus toxin, the values were <1.0 IU/ml, 1.0 to 10.0 IU/ml, and >10.0 IU/ml. A level of >1.0 IU/ml is indicative of immune protection. For IgG anti-diphtheria toxin, the values were <0.1 IU/ml, 0.1 to 2.0 IU/ml, and >2.0 IU/ml. A level of >0.1 IU/ml is regarded as the threshold for immune protection. For IgG anti-PT, the values were <50 EU/ml, 50 to 100 EU/ml, and >100 EU/ml. Since no accepted levels of protection for antibodies for pertussis exist, a level of <50 EU/ml was regarded as probably indicative of no recent contact, a level between 50 EU/ml and 100 EU/ml was regarded as indeterminate, and a level of >100 EU/ml was regarded as an indicator of recent contact (1).

RESULTS

Antibodies to tetanus toxin.The concentrations of antibodies measured by the multiplex assay were plotted in scatter plots against the ELISA values (Fig. 1). A regression coefficient of 0.910 was found compared to the reference ELISA. The y intercept of the curve was −0.12 IU/ml, and the slope was 1.17. Compared to the reference ELISA, the regression coefficients of the other ELISA systems were 0.935 (Euroimmun) and 0.832 (Virotech), respectively. The intercepts of the other ELISA systems were −0.31 (Euroimmun) and −0.32 (Virotech), respectively, and the slopes of the regression curves were 1.21 (Euroimmun) and 0.964 (Virotech), respectively. Another scatter plot (Fig. 2) shows the values of the multianalyte system and two other commercial ELISAs compared to the reference ELISA. Values are displayed in four different ranges: <1.0 IU/ml, 1.0 to 5.0 IU/ml, 5.0 to 10.0 IU/ml, and >10 IU/ml. It is obvious that the overlapping of ranges was similar between all systems compared. A Bland-Altman plot of the bias between the reference ELISA and the multianalyte system (Fig. 3) shows no relevant bias between both methods.

FIG. 1.
  • Open in new tab
  • Download powerpoint
FIG. 1.

Scatter plot of values measured by the reference IgG anti-tetanus toxin ELISA versus those measured by the multianalyte assay (IU/ml).

FIG. 2.
  • Open in new tab
  • Download powerpoint
FIG. 2.

Plot of values in different ranges of IgG anti-tetanus toxin as measured by the multianalyte system (Multi) and two commercially available ELISAs compared to the reference ELISA. Symbols: ▪, <1.0 IU/ml; ○, 1.0 to 5.0 IU/ml; •, 5.0 to 10.0 IU/ml; ▵, >10 IU/ml. The concordance between the reference ELISA, two other ELISA systems, and the multianalyte system is demonstrated.

FIG. 3.
  • Open in new tab
  • Download powerpoint
FIG. 3.

Anti-tetanus toxin. A Bland-Altman plot of differences between reference ELISA and multianalyte system is shown.

Table 1 compares the concordance in which three diagnostically relevant cutoffs were used. These were <0.1 IU/ml, 0.1 to 1.0 IU/ml, 1.0 to 10.0 IU/ml, and >10 IU/ml. The systems agreed by 96, 72, and 94%, respectively (Tables 1 to 3). In the medium range (1.0 to 10 IU/ml) the multiplex assay showed higher values (19% were >10 IU/ml). When the multianalyte assay was compared to the Virotech assay, both tests agreed by 96, 74, and 63%, and with the Euroimmun ELISA the concordance was 100, 85, and 87%.

View this table:
  • View inline
  • View popup
TABLE 1.

Concordance of ranges for IgG anti-tetanus toxin antibodies (reference ELISA versus multianalyte assay)

Antibodies to diphtheria toxin.The concentrations of antibodies measured by the multiplex assay were plotted in scatter plots against the ELISA values (Fig. 4). A linear regression of 0.938 was found compared to the reference ELISA. The y intercept of the curve was −0.11 IU/ml, and the slope was 1.19. Compared to the reference ELISA, the regression coefficients of the other ELISA systems were 0.934 (Euroimmun) and 0.916 (Virotech), respectively. The intercepts of the other ELISA systems were −0.67 (Euroimmun) and −0.19 (Virotech), respectively, and the slopes of the regression curves were 1.14 (Euroimmun) and 0.963 (Virotech), respectively.

FIG. 4.
  • Open in new tab
  • Download powerpoint
FIG. 4.

Scatter plot of values measured by the reference IgG anti-diphtheria toxin ELISA versus those measured by the multianalyte assay (IU/ml).

Figure 5 shows the values of the multianalyte system and two other commercial ELISAs compared to the reference ELISA. Values are displayed in four different ranges: <0.1 IU/ml, 0.1 to 1.0 IU/ml, 1.0 to 2.0 IU/ml, and >2.0 IU/ml. It is obvious that the overlapping of ranges was comparable between the systems. The Virotech ELISA had a lower level of detection of 0.1 IU/ml. A Bland-Altman plot of the bias between the reference ELISA and the multianalyte system (Fig. 6) shows no relevant bias between the two methods.

FIG. 5.
  • Open in new tab
  • Download powerpoint
FIG. 5.

Plot of values in different ranges of IgG anti-diphtheria toxin as measured by the multianalyte system (Multi) and two commercially available ELISAs compared to the reference ELISA. Symbols: ▪, <0.1 IU/ml; ○, 0.1 to 1.0 IU/ml; •, 1.0 to 2.0 IU/ml; ▵, >2.0 IU/ml.

FIG. 6.
  • Open in new tab
  • Download powerpoint
FIG. 6.

Anti-diphtheria toxin. A Bland-Altman plot of differences between reference ELISA and multianalyte system is shown.

Table 2 compares the diagnostically relevant cutoffs. For anti-diphtheria toxin, these were <0.1 IU/ml, 0.1 to 2.0 IU/ml, and >2 IU/ml. The systems agreed by 100, 83, and 94%, respectively. When the multiplex assay was compared to the Virotech assay, both tests agreed by 92, 68, and 75%, and with the Euroimmun ELISA the concordance was 92, 60, and 5%.

View this table:
  • View inline
  • View popup
TABLE 2.

Concordance of ranges for IgG anti-diphtheria toxin antibodies (reference ELISA versus multianalyte assay)

Antibodies to pertussis antigens (IgG anti-PT).The concentrations of antibodies measured by the multianalyte assay were plotted in scatter plots against the ELISA values (Fig. 7). A linear regression of 0.905 was found compared to the reference ELISA. The y intercept of the curve was 0.68 EU/ml, and the slope was 0.81. Compared to the reference ELISA, the regression coefficients of the other ELISA systems were 0.869 (Virion) and 0.816 (Virotech), respectively. The intercepts of the other ELISA systems were 0.40 (Virion) and 0.36 (Virotech), recepectively, and the slopes of the regression curves were 0.87 (Virion) and 0.42 (Virotech), respectively.

FIG. 7.
  • Open in new tab
  • Download powerpoint
FIG. 7.

Scatter plot of values of the reference “in-house” IgG anti-PT ELISA versus those measured by the multianalyte assay (EU/ml).

Figure 8 shows the values of the multianalyte system and two other commercial ELISAs compared to the reference ELISA. Values are displayed in three different ranges: <50 EU/ml, 50 to 100 EU/ml, and >100 EU/ml. The Virotech ELISA uses a different unitage, which is shown on the right y axis of the figure. A Bland-Altman plot of the bias between the reference ELISA and the multianalyte system (Fig. 9) showed that the reference ELISA measured systematically lower levels, when sera contained medium to high concentrations of anti-PT.

FIG. 8.
  • Open in new tab
  • Download powerpoint
FIG. 8.

Plot of values in different ranges of IgG anti-PT as measured by the multianalyte system (Multi) and two commercially available ELISAs compared to the reference ELISA. Symbols: ▪, <50 EU/ml; ○, 50 to 100 EU/ml; ▵, >100 EU/ml.

FIG. 9.
  • Open in new tab
  • Download powerpoint
FIG. 9.

Anti-PT. A Bland-Altman plot of differences between reference ELISA and multianalyte system is shown.

Table 3 also displays possible diagnostically relevant cutoffs for anti-PT. These were <50 EU/ml, 50 to 100 EU/ml, and >100 EU/ml. The systems agreed by 82, 9, and 98% (Tables 1 to 3). As also shown in Table 3, the multiplex assay identified 91% of the sera which were between 50 and 100 EU/ml in the reference assay as being >100 EU/ml. When we compared the multiplex assay to the Serion ELISA, both tests agreed by 87, 30, and 94%, respectively. The Virotech ELISA measures in different units, so a comparison could not be made.

View this table:
  • View inline
  • View popup
TABLE 3.

Concordance of ranges for IgG anti-PT antibodies (reference ELISA versus multianalyte assay)

All methods showed a 100% agreement for the serum pairs. The increase in signal after vaccination was significant in all ELISAs and the multianalyte method.

DISCUSSION

Antibodies to tetanus toxin as a result of immunization have been measured primarily by mouse neutralization tests. ELISA formats have reproducibly yielded comparable results, and they are now widely used to measure the vaccination response (8). Primarily derived from the mouse model, protective cutoffs for antibodies have been established, and these are now widely used in seroepidemiological studies to measure the population immunity to this antigen.

Similarly, antibodies to diphtheria toxin were primarily measured with an animal system, i.e., a guinea pig intradermal test. In addition to ELISA systems, which are used for routine purposes, the antibody response to vaccination can also reliably be measured by a Vero cell assay (9). Similar to tetanus toxin, protective thresholds have been defined that should guarantee protection against the toxin with a high level of probability.

ELISA systems that measure antibodies to tetanus and diphtheria toxin are commercially available; these assays have to be licensed by the FDA in the United States, and they must comply with the in vitro diagnostic directive of the European Union. Methods for measuring antibodies to B. pertussis antigens such as PT or FHA are less well standardized. Commercially available tests in the European Union show variable results (4), and no FDA-approved test is available in the United States. In contrast to antibodies to tetanus or diphtheria toxin, no serological correlate for protective immunity exists (1, 2).

Although ELISA systems are well characterized and rather reliable, they require an aliquot of a patient sample for every test, and the cost for labor and materials increases with the number of analyses.

A microsphere-based flow cytometry technique circumvents these issues by analyzing several parameters in one sample simultaneously (5, 6, 7, 10). Flow cytometers are widely used in laboratories, and thus this assay system can use existing equipment. In a multiplex bead assay, serum antibodies are in closer contact with the antigens bound to the beads, making the test faster. Compared to single ELISAs, the overall turnover time is much shorter and less sample volume is needed. The simultaneous analysis of all samples allows for the simultaneous validation of results.

Among various other applications of this technique, methods have been described to measure antibodies to diphtheria and tetanus toxin and Haemophilus influenzae type b capsular polysaccharide (5) and to PT and FHA of B. pertussis (6). These studies used other antigen sources and compositions and also different ELISA comparators. Given these caveats, the multiplex bead assay compared well in our study and in a previous study (5) with single ELISAs for measuring antibodies to tetanus and diphtheria toxoids. The multiplex assay used previously (6) was meant as a diagnostic tool and thus measured IgG and IgA antibodies to PT and FHA. With respect to IgG anti-PT, the concordance between the ELISA and the multianalyte system was 82% for concentrations of <50 EU/ml, which was comparable (6). Sera with high levels of anti-PT were measured mostly lower with the reference ELISA. However, the multianalyte system detected all titer increases after vaccination that were detected by single-antigen ELISAs and thus seems well suited for measuring vaccine responses.

In conclusion, the multiplex assay tested here was a robust system, providing rapid and accurate results. It showed good concordance with various single-antigen ELISAs and performed well in measuring the immune response after vaccination. It thus offers an alternative to ELISA systems for measuring vaccine responses in routine and research laboratories.

FOOTNOTES

    • Received 1 June 2007.
    • Returned for modification 1 October 2007.
    • Accepted 22 February 2008.
  • Copyright © 2008 American Society for Microbiology

REFERENCES

  1. 1.↵
    Baughman, A. L., K. M. Bisgard, K. M. Edwards, D. Guris, M. D. Decker, K. Holland, B. D. Meade, and F. Lynn. 2004. Establishment of diagnostic cutoff points for levels of serum antibodies to pertussis toxin, filamentous hemagglutinin and fimbriae in adolescents and adults in the United States. Clin. Diagn. Lab. Immunol.11:1045-1053.
    OpenUrlCrossRefPubMed
  2. 2.↵
    Edwards, K. 2004. Pertussis Vaccine, p. 471-528. In S. A. Plotkin and W. A. Orenstein (ed.), Vaccines, 4th ed. Saunders, Philadelphia, PA.
  3. 3.↵
    Knuf, M., F. Zepp, C. Meyer, E. Grzgegowski, J. Wolter, M. Riffelmann, and C. H. Wirsing von Koenig. 2006. Immunogenicity of a single dose of reduced-antigen acellular pertussis vaccine in a non-vaccinated adolescent population. Vaccine24:2043-2048.
    OpenUrlCrossRefPubMedWeb of Science
  4. 4.↵
    Koesters, K., M. Riffelmann, B. Dohrn, and C. H. Wirsing von Koenig. 2000. Comparison of six commercially available tests for detecting antibodies to Bordetella antigens. Clin. Lab. Diagn. Immunol.7:422-426.
    OpenUrlAbstract/FREE Full Text
  5. 5.↵
    Pickering, J. W., T. B. Martins, M. C. Schroder, and H. R. Hill. 2002. Comparison of a multiplex flow cytometric assay with enzyme-linked immunosorbent assay for quantification of antibodies to tetanus, diphtheria, and Haemophilus influenzae type b. Clin. Diagn. Lab. Immunol.9:872-876.
    OpenUrlCrossRefPubMed
  6. 6.↵
    Prince, H. E., M. Lapé-Nixon, and J. Matud. 2006. Evaluation of a tetraplex microsphere assay for Bordetella pertussis antibodies. Clin. Vaccine Immunol.13:266-270.
    OpenUrlAbstract/FREE Full Text
  7. 7.↵
    Tripatzis, I. 1983. German patent DE3322373C2. Testmittel und Verfahren zum Nachweis von Antigenen und/oder Antikörpern.
  8. 8.↵
    Wassilak S. G. F., T. V. Murphy, M. H. Roper, and W. A. Orenstein. 2004. Tetanus toxoid, p. 742-782. In S. A. Plotkin and W. A. Orenstein (ed.), Vaccines, 4th ed. Saunders, Philadelphia, PA.
  9. 9.↵
    Wharton, M., and C. R. Vitek. 2004. Diphtheria toxoid, p. 211-228. In S. A. Plotkin and W. A. Orenstein (ed.), Vaccines, 4th ed. Saunders, Philadelphia, PA.
  10. 10.↵
    Wilson, M. R., and J. S. Wotherspoon. 1988. A new microsphere-based immunofluorescence assay using flow cytometry. J. Immunol. Methods107:225-230.
    OpenUrlCrossRefPubMedWeb of Science
  11. 11.↵
    Wirsing von König, C. H., D. Gounis, H. J. Schmitt, S. Laukamp, and H. Bogaerts. 1999. Single sample serological diagnosis of pertussis. Eur. J. Clin. Microbiol. Infect. Dis.18:341-345.
    OpenUrlCrossRefPubMed
View Abstract
PreviousNext
Back to top
Download PDF
Citation Tools
Measuring Immunoglobulin G Antibodies to Tetanus Toxin, Diphtheria Toxin, and Pertussis Toxin with Single-Antigen Enzyme-Linked Immunosorbent Assays and a Bead-Based Multiplex Assay
Sabine Reder, Marion Riffelmann, Christian Becker, Carl Heinz Wirsing von König
Clinical and Vaccine Immunology May 2008, 15 (5) 744-749; DOI: 10.1128/CVI.00225-07

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
Print

Email

Thank you for sharing this Clinical and Vaccine Immunology article.

NOTE: We request your email address only to inform the recipient that it was you who recommended this article, and that it is not junk mail. We do not retain these email addresses.

Enter multiple addresses on separate lines or separate them with commas.
Measuring Immunoglobulin G Antibodies to Tetanus Toxin, Diphtheria Toxin, and Pertussis Toxin with Single-Antigen Enzyme-Linked Immunosorbent Assays and a Bead-Based Multiplex Assay
(Your Name) has forwarded a page to you from Clinical and Vaccine Immunology
(Your Name) thought you would be interested in this article in Clinical and Vaccine Immunology.
CAPTCHA
This question is for testing whether or not you are a human visitor and to prevent automated spam submissions.
Share
Measuring Immunoglobulin G Antibodies to Tetanus Toxin, Diphtheria Toxin, and Pertussis Toxin with Single-Antigen Enzyme-Linked Immunosorbent Assays and a Bead-Based Multiplex Assay
Sabine Reder, Marion Riffelmann, Christian Becker, Carl Heinz Wirsing von König
Clinical and Vaccine Immunology May 2008, 15 (5) 744-749; DOI: 10.1128/CVI.00225-07
del.icio.us logo Digg logo Reddit logo Twitter logo CiteULike logo Facebook logo Google logo Mendeley logo
  • Top
  • Article
    • ABSTRACT
    • MATERIALS AND METHODS
    • RESULTS
    • DISCUSSION
    • FOOTNOTES
    • REFERENCES
  • Figures & Data
  • Info & Metrics
  • PDF

KEYWORDS

Antibodies, Bacterial
Enzyme-Linked Immunosorbent Assay
Immunoassay
Immunoglobulin G

Related Articles

Cited By...

About

  • About CVI
  • For Librarians
  • For Advertisers
  • FAQ
  • Permissions
  • Journal Announcements

Authors

  • Submit a Manuscript to mSphere

ASM Journals

ASM journals are the most prominent publications in the field, delivering up-to-date and authoritative coverage of both basic and clinical microbiology.

About ASM | Contact Us | Press Room

 

ASM is a member of

Scientific Society Publisher Alliance

 

American Society for Microbiology
1752 N St. NW
Washington, DC 20036
Phone: (202) 737-3600

Copyright © 2021 American Society for Microbiology | Privacy Policy | Website feedback

Print ISSN: 1556-6811; Online ISSN: 1556-679X