Clinical and Vaccine Immunology, August 2008, p. 1304-1306, Vol. 15, No. 8
1071-412X/08/$08.00+0 doi:10.1128/CVI.00157-08
Copyright © 2008, American Society for Microbiology. All Rights Reserved.

Focus Diagnostics Inc., Cypress, California
Received 2 May 2008/ Returned for modification 29 May 2008/ Accepted 9 June 2008
|
|
|---|
|
|
|---|
A major laboratory tool used in diagnosing dengue virus infections is measurement of the level of dengue virus immunoglobulin M (IgM) in serum (5, 6). Most dengue virus IgM tests utilize the µ-capture enzyme-linked immunosorbent assay (ELISA) format (1, 3, 4, 9, 14), which employs capture wells coated with anti-human IgM, inactivated dengue virus antigen, and enzyme-conjugated murine anti-flavivirus monoclonal antibody (reporter reagent). Because captured IgM with heterophilic antibody activity may yield false-positive results by directly binding the reporter reagent (8, 10), we employ a two-step testing algorithm for dengue virus IgM detection. Sera are first screened using a µ-capture ELISA (one well per sample); samples with equivocal or positive results are then retested using a background subtraction (BS) modification of this same ELISA (two wells per sample), designed to identify false-positive screening assay reactivity due to heterophilic antibodies (8, 12, 13). Because the expected screening ELISA reactivity rate is <50%, using this algorithm consumes fewer capture wells and less reporter reagent than testing all samples using the BS ELISA.
In conjunction with a surge in the number of dengue cases in Mexico and the Caribbean during the summer and fall of 2007 (2, 11), our reference laboratory tested 2,692 consecutive sera for dengue virus IgM. We capitalized on this large number of dengue virus IgM results to evaluate the efficiency of our testing algorithm. Specifically, we asked if there is a screening ELISA value above which false-positive results are not observed; if such a value exists, then sera with screening ELISA results above this value can be interpreted as dengue virus IgM positive without performing the BS ELISA, thus improving the algorithm's efficiency.
Sera submitted for dengue virus antibody testing were evaluated using an in-house-developed dengue virus IgM screening ELISA; this assay utilizes many of the components found in the dengue virus IgM ELISA kit manufactured by the diagnostic products division of Focus Diagnostics, Cypress, CA (the kit is for research use only in the United States) (1, 9). Each assay included negative control serum, positive control serum, and calibrator serum (Focus Diagnostics). Control, calibrator, and patient sera were diluted 1:101 in specimen diluent (Focus Diagnostics), and 0.1 ml of diluted sample was added to an assigned microtiter well coated with rabbit anti-human IgM (Focus Diagnostics). After incubation for 1 h at room temperature (RT), the wells were washed three times and then received inactivated dengue virus antigen (containing all four dengue virus serotypes [Focus Diagnostics]). After incubation for 2 h at RT and washing, the wells received horseradish peroxidase-conjugated 6B6C anti-flavivirus monoclonal antibody (Focus Diagnostics). After incubation for 30 min at RT and washing, the wells received tetramethylbenzidine (enhanced K-blue; Neogen Corp., Lexington, KY); after 10 min, the color reaction was stopped by adding sulfuric acid (Ricca Chemicals, Arlington, TX). Absorbance at 450 nm was measured using an ELISA reader (BioTek, Winooski, VT). The results were expressed as an index, calculated by dividing the specimen absorbance value by the calibrator absorbance value; indices of <0.90 were considered negative, indices from 0.90 to 1.10 were considered equivocal, and indices of >1.10 were considered positive.
Sera with equivocal or positive results in the dengue virus IgM screening ELISA were tested using the BS ELISA (8, 12). In this modified screening ELISA, all control, calibrator, and patient sera were added to two capture wells. After incubation for 1 h at RT and washing, dengue virus antigen was added to one well and specimen diluent was added to the other well. The assay was then finished per the screening ELISA procedure. For each specimen (including controls and calibrator), the absorbance value of the well receiving specimen diluent was subtracted from the absorbance value of the well receiving dengue virus antigen. This corrected absorbance value was then used to calculate the index. As with the screening ELISA, BS ELISA indices of <0.90 were considered negative, indices from 0.90 to 1.10 were considered equivocal, and indices >1.10 were considered positive.
Table 1 presents representative results for the screening ELISA and BS ELISA, demonstrating how index values were calculated. Data for calibrator sera, controls, one screen false-positive patient sample, and two true-positive patient samples are shown.
|
View this table: [in a new window] |
TABLE 1. Results from a representative dengue virus IgM screening ELISA and subsequent BS ELISA
|
![]() View larger version (11K): [in a new window] |
FIG. 1. Summary of dengue virus IgM screening ELISA and BS ELISA results for 2,962 consecutive sera submitted for dengue virus IgM testing. For both assays, indices of <0.90 were considered negative, indices from 0.90 to 1.10 were considered equivocal, and indices of >1.10 were considered positive.
|
|
View this table: [in a new window] |
TABLE 2. Relationship of dengue virus IgM BS ELISA results to screening ELISA index values
|
The BS approach for identifying false-positive reactivity is routinely applied to other screening µ-capture ELISA systems besides dengue virus IgM (e.g., West Nile virus IgM) (8, 13). Thus, it stands to reason that these other screening assays, like the dengue virus IgM assay, may also have a characteristic high index cut point above which BS ELISA performance is not necessary. This cut point will undoubtedly vary among different assays, depending on the absorbance value of the screening ELISA calibrator and the dynamic range of the assay. Each laboratory must therefore define its own reflex testing algorithms for analytes measured by µ-capture ELISA.