Clinical and Vaccine Immunology, July 2008, p. 1054-1059, Vol. 15, No. 7
1071-412X/08/$08.00+0 doi:10.1128/CVI.00008-08
Copyright © 2008, American Society for Microbiology. All Rights Reserved.

Mayo Vaccine Research Group, Mayo Clinic College of Medicine, Rochester, Minnesota 55905,1 Department of Health Sciences Research, Mayo Clinic College of Medicine, Rochester, Minnesota 55905,2 Program in Translational Immunovirology and Biodefense, Mayo Clinic College of Medicine, Rochester, Minnesota 559053
Received 21 December 2007/ Returned for modification 6 February 2008/ Accepted 25 April 2008
|
|
|---|
|
|
|---|
Neutralizing and protective antibodies are directed against the two surface MV glycoproteins, the hemagglutinin (H) and fusion (F) proteins, and are sufficient to provide protection (4). Therefore, diagnostic measures of sufficient levels of functional neutralizing MV-specific antibodies correlate with protection. Neutralizing antibodies are operationally defined by a standard plaque reduction neutralization (PRN) test as antibodies that prevent a cytopathic effect and formation of plaques, using a laboratory MV strain (Edmonston) on Vero cells (4). The PRN assay, an enhanced version of the neutralization test, is widely accepted as the "gold standard" in MV serology and measures the serum dilution capable of preventing 50% of plaque formation by MV (PRN titer; 50% neutralizing dose [ND50]) (1). However, the classical PRN assays are slow (4 to 7 days), labor-intensive, and require large volumes of test components (since they are conducted in 24-well and 12-well plates using semisolid cell overlay and staining procedures) and are impractical for large numbers of samples. The availability of a relatively simple, more rapid, reproducible, and sensitive assay for MV neutralizing antibodies, applicable to larger-scale studies, would be of considerable interest for vaccine-related research.
For this reason, we have developed and standardized a simplified, highly sensitive fluorescence-based plaque reduction microneutralization (PRMN) assay, using recombinant MV based on the Edmonston B strain, engineered to express enhanced green fluorescent protein, MVeGFP (9, 19). The assay was validated using the second and third World Health Organization (WHO) international anti-MV standards, one in-house reference serum (calibrated against the WHO standard), and three groups of low, moderate, and high MV vaccine responders sera with good concordance in antibody titers between PRMN and the Dade Behring immunoglobulin G (IgG) enzyme immunoassay (EIA).
|
|
|---|
MV IgG EIA.
Quantitative levels of MV-specific IgG antibody titers for all serum specimens were determined using the Dade Behring anti-MV IgG EIA (Dade Behring, Marburg, Germany; sensitivity, 99.6%; specificity, 100%), according to the manufacturer's instructions as described previously (8). The assay is based on cell culture-grown native virus antigen (simian kidney cells infected with MV) and control antigen. Serum samples (dilution, 1:231) were assayed in duplicate, and the difference in mean absorbance between the test antigen and the control antigen for each sample was calculated and multiplied by a correction factor according to the Dade Behring assay correction protocol to give the corrected change in absorbance (
A). MV-specific IgG antibody levels (in mIU/ml) were calculated from the antilog of the following formula: log10 mIU/ml =
x
Aβ, where
and β are lot-dependent constants. Negative for anti-MV antibody was defined as a
A of less than 0.1 (150 mIU/ml), and positive for anti-MV antibodies was defined as a
A of greater than 0.2. Test samples yielding a result within the above range (0.1
A
0.2) were considered equivocal. The coefficient of variation for this assay in our laboratory was 3.8% as determined in our previous studies (8).
PRMN assay.
The PRMN assay was based on the classical PRN assay (1, 6, 7, 18) and was performed in 96-well flat-bottom plates. Thirty test sera (10 samples with the highest IgG antibody levels [geometric mean
A, 1.403; geometric mean titer [GMT], 8,610 mIU/ml], 10 samples with the lowest IgG antibody levels [geometric mean
A, 0.096; GMT, 150 mIU/ml], and a random sample of 10 sera in the middle 50% of MV IgG antibody levels [geometric mean
A, 0.595; GMT, 1,746 mIU/ml]) were heat inactivated (56°C; 30 min) and assayed 3 times (or up to 20 times for the standards) on separate days for neutralizing antibodies by PRMN. Serum samples were diluted fourfold from 1:4 to 1:4,096 (four replicates for each dilution) in Opti-MEM I (Gibco, Invitrogen Corporation, Carlsbad, CA), except for the second and the third WHO international anti-MV standards (5 IU, National Institute for Biological Standards and Control [NIBSC] code no. 66/202, and 3 IU, NIBSC code no. 97/648; WHO International Laboratory for Biological Standards, NIBSC, Potters Bar, Hertfordshire, United Kingdom), which were diluted fourfold from 1:16 to 1:16,384. The diluted sera were mixed with equal volumes of low-passage challenge virus MVeGFP (9, 19) (final dilutions, 1:8 to 1:8,192 for all sera and 1:32 to 1:32,768 for the WHO standards) and incubated for 1 h at 37°C. A standard inoculum of challenge virus was used in Opti-MEM at a dilution adjusted to yield 15 to 30 PFU per well in the control wells with virus only. Serum-virus mixtures (50 µl) were transferred to a new 96-well plate, mixed with an equal volume of Vero cell suspension (1.5 x 104/well; Vero passage level, 20 to 40 passages), and incubated for 2 h at 37°C, after which 50 µl of Dulbecco's modified Eagle's medium (Gibco, Invitrogen Corporation, Carlsbad, CA) containing 10% fetal bovine serum (HyClone, Logan, UT) was added. The plates were incubated for 48 h at 37°C under 5% CO2, and the numbers of brightly fluorescent green plaques were counted under UV light using a fluorescence microscope (Nikon Eclipse TE300; Nikon Instruments Inc., Melville, NY). For automatic readout, 48 h after assay setup, the cells were lysed with 4% Igepal CA630 (Sigma Aldrich) in Dulbecco's modified Eagle's medium-10% fetal bovine serum (incubation, 1 h at 37°C), and GFP fluorescence was measured using a fluorescence reader (Fluoroscan Ascent; Thermo Electron Corporation, Vantaa, Finland) at 485-nm excitation and 527-nm emission wavelengths. The PRMN assay controls included the second and third WHO international anti-MV antibody standards, containing 5,000 mIU/ml and 3,000 mIU/ml, respectively, and one in-house reference serum, calibrated against the second WHO standard. The PRMN values (ND50) were calculated by Karber's formula (2) or using nonlinear regression analysis (Microsoft Excel 2004 version 11.3.7) to calculate the titer that gave a 50% reduction in GFP fluorescence compared to a virus-only control. PRN values of <8 were considered to be negative. PRN values of >120 were thought to indicate protection from symptomatic disease (5), whereas values of >800 (20) or 1,052 (5) indicated sterile immunity and protection from infection with no antibody boosting after MV exposure. The use of the second WHO international anti-MV antibody standard enabled quantitative ND50 values to be transformed into mIU/ml, as described previously (6). A test limit in terms of mIU/ml was determined for each assay (6). Test sera with reactivities greater than the test limit (corresponding to a PRMN value of 8) were considered PRMN positive.
Statistical methods. Data were descriptively summarized using frequencies and percents for categorical variables and means and ranges for continuous variables. Plots of antibody levels (in mIU/ml or both ND50 and mIU/ml) suggested moderate to high levels of data skewness, so all calculations of summary statistics and all formal statistical analyses for these measures were carried out using log-transformed values. The resulting mean values were then back transformed and thus expressed in their original sampling units, effectively yielding GMT values. We determined the reproducibility of the PRMN assay using Shewhart control charts, plotting replicate values of the second WHO standard and third WHO standard over time. Control limits, depicting the means ± 2 standard deviations (SD), were superimposed on each of the charts. Since EIAs and PRMN (PRN) assays are not expected to give equivalent quantitative values (6, 18), we did not statistically compare the EIA and PRMN values but rather looked for correlations between them. We visually displayed associations between neutralizing-antibody levels and EIA values using scatter plots. Formal associations between the two sets of antibody measures were examined in two ways. Primary analyses used general linear mixed models, accounting for the repeated measures of neutralizing-antibody levels within an individual. Subject-specific effects were modeled using a random intercept term, and the resulting variance component was used to calculate an intraclass correlation coefficient, which can be interpreted as the within-person correlation between the neutralizing antibody and the EIA values. Secondary analyses used simple Pearson correlation coefficients to examine associations between EIA values and neutralizing-antibody responses. In contrast to the intraclass correlations, these analyses were performed with one observation per subject, using the mean of the three replicate PRMN values. Associations were examined both overall and within each of the three EIA-defined sampling categories. All statistical tests were two sided, and all analyses were carried out using the SAS system (SAS Institute, Inc., Cary, NC).
|
|
|---|
A > 0.2), while the remaining 14% were negative or equivocal for MV-specific antibody (
A
0.2). For the PRMN validation studies, we used three EIA sampling groups, defined by the EIA antibody titer in mIU/ml: 10 sera with the lowest IgG antibody level comprised the low-antibody-level group (EIA GMT, 150 mIU/ml; geometric mean
A, 0.096), 10 sera with the highest IgG antibody levels comprised the high-antibody-level group (EIA GMT, 8,610 mIU/ml; geometric mean
A, 1.403), and a random sample of 10 sera in the middle 50% of MV IgG antibody levels comprised the moderate-antibody-level group (EIA GMT, 1,746 mIU/ml; geometric mean
A, 0.595). The group with the lowest IgG antibody levels included three EIA-negative sera (
A < 0.1) and seven EIA-equivocal sera (
A
0.2). Establishment of PRMN. To establish and characterize the performance of the PRMN assay, we initially used the two available WHO international anti-MV standards with assigned values (second standard 66/202 to 5 IU and third standard 97/648 to 3 IU) and known performance in the classical PRN assay. In addition, we also used a diluted WHO standard (66/202) working stock of 200 mIU/ml as described by Ratnam et al. (18) and one in-house high-titer reference serum. All sera were tested against the second WHO anti-MV serum 66/202 as a standard, which allowed us to transform all ND50 values into mIU/ml values. After identifying the dynamic range for these products, we conducted multiple PRMN assays using a standard operating procedure over a period of several months. The reproducibility and variability of the PRMN assay were determined by the generation of quality control charts in which the mean ± 2 SD limits for the standards were determined (Fig. 1). The reproducibility of our assay was found to be lower than or comparable to that shown for the classical PRN assay (7) and for other neutralization assays, based on reporter gene expression (14). The geometric mean PRMN titer (ND50) of the second WHO standard 66/202 identified in 20 repeated assays was 2,061 versus a mean ND50 value of 2,350, found for the same standard by the classical PRN assay (7). The performance of the PRMN assay on low (200-mIU/ml standard by PRN)- and high (3,000-mIU/ml third WHO standard by PRN)-antibody-level standards was also examined. The geometric mean PRMN titer of the third WHO standard 97/648 for 20 repeated tests was 1,107, and the geometric mean calculated value was 2,948 mIU/ml (the assigned value for this standard is 3,000 mIU/ml as determined by the classical PRN assay). The second WHO diluted working stock of 200 mIU/ml was tested 10 times, and the geometric mean PRMN titer (ND50) was 82 versus a PRN titer of 94, found for the same standard by Ratnam et al. (18), while the geometric mean calculated value in mIU/ml was 211. In a set of repeated assays, we were also able to assess the performance of our in-house reference serum using our PRMN assay with respect to its future use in this assay as a standard. The geometric mean PRMN titer (ND50) of this serum was 1,207, and the calculated GMT value was 2,999 mIU/ml. Thus, the in-house reference serum was assigned a value of 3,000 mIU/ml for future studies. In 10 repeated assays of the third WHO standard 97/648, we compared the visual and the automatic readouts. The geometric mean ND50 value based on visual plaque counting was 927, while the geometric mean ND50 value calculated from the measured GFP fluorescence was 980, and there was no significant difference between the mean values from visual and automatic readouts (P = 0.45; t test on log-transformed values).
![]() View larger version (17K): [in a new window] |
FIG. 1. WHO anti-MV standard control charts in PRMN. (A) Second WHO anti-MV standard 66/202 control chart showing PRMN ND50 values in 20 repeated assays. (B and C) Third WHO anti-MV standard 97/648 control charts showing PRMN ND50 values (B) and PRMN mIU/ml values (C) in 20 repeated assays. The values are plotted on a log scale. The dashed horizontal lines depict the sample mean ± 2 SD limits.
|
![]() View larger version (33K): [in a new window] |
FIG. 2. PRMN, using MVeGFP, of representative low-, moderate-, and high-antibody-titer sera. The photographs of brightly fluorescent plaques were taken 48 h after PRMN setup under UV light on a fluorescence microscope (40x magnification). Plaques in the different dilutions were counted and compared with the virus-only control to calculate the PRMN titer using Karber's formula (2).
|
|
View this table: [in a new window] |
TABLE 1. Antibody titers and 50% neutralizing doses of test sera by EIA and PRMN
|
![]() View larger version (18K): [in a new window] |
FIG. 3. Association between MV neutralizing-antibody and EIA levels. (A) Values are plotted on a log scale. Each individual is represented three times, once per measure of neutralizing antibodies. The dashed lines separate the three sampling strata as defined by EIA values. The solid line is the 45° line of equivalence. Most values fall to the right of the line, indicating that on average EIA values are larger than neutralizing-antibody values. (B) Correlations between EIA and neutralizing-antibody response measures, overall and by EIA-defined subgroups. Interaction correlations are for three neutralizing-antibody measures per individual. Pearson correlations are based on one observation per individual, using the mean of the three neutralizing-antibody measures. Overall results for the Pearson correlation should be interpreted with caution, as they are more heavily affected by the sampling design than the intraclass correlations.
|
![]() View larger version (17K): [in a new window] |
FIG. 4. Relevance of PRMN ND50 values to protective antibody levels. PRMN values of low-, moderate-, and high-antibody-titer sera (30 observations per category) were plotted on a log scale, and the antibody levels relevant to protection are shown as dotted lines: 120 (5), threshold for protection against symptomatic disease; 800 (20) or 1,052 (5), threshold for protection against infection and antibody boosting after MV exposure.
|
|
|
|---|
The performance of the two WHO international anti-MV standards in PRMN gave us the opportunity to develop and validate the test, to transform all values into mIU/ml, and to compare our results with those of the classical PRN assay, since the WHO standards values were established and designed for use mainly in the PRN assay. The PRMN assay requires substantially less incubation time (2 days versus 4 to 7 days) and fewer test components (no overlay medium and no staining and fixation reagents) than the classical PRN assay and eliminates the complicated steps of semisolid cell overlays, fixation, and staining. Moreover, the GFP expression greatly facilitates and improves the test readout. We have demonstrated that the GFP-based PRMN assay is rapid, reproducible, and sensitive and produces results similar to those obtained by the classical PRN assay. The PRMN assay values from 20 repeated assays were remarkably similar to those reported in the literature (7, 18) or assigned to the WHO standards by the classical PRN assay. The new assay is basically a PRN assay with visual (plaque enumeration) and/or automatic readout, which renders results consistent with those in the classical PRN assay. This is of utmost importance, since the reported PRN antibody levels relevant to protection could be directly applied to the interpretation of PRMN assay results. However, an important limitation of a PRMN assay with visual readout is human subjectivity. While the fluorescence reader-based assay offers clear advantages in time and human resources, it does require specific instrumentation. The potential sources of error with the latter method, such as high background fluorescence, edge effects, and sample fluorescence/quenching, should be addressed prior to implementation.
Several EIA kits for anti-MV antibodies are available commercially, among which the Dade Behring EIA kit was previously found to perform better in comparison with other commercial EIA kits based on a PRN titer of 120 as a cutoff (12, 18). While these kits are easy to use, they are generally not sufficiently sensitive to detect low levels of antibodies and are therefore not optimal for seroprevalence surveys (11). To validate our new assay, we used low, moderate, and high MV vaccine responders sera as determined by the Dade Behring EIA. The new assay was sensitive enough to detect both low-titer and high-titer neutralizing antibodies in clinical serum samples. Similar to previous reports for PRN (6, 11), we found that the PRMN assay showed much better sensitivity than the Dade Behring EIA kits and was able to detect low levels of anti-MV antibodies in samples that gave negative/equivocal results using the EIA. Analyzing the associations between MV PRMN antibody levels in mIU/ml and EIA antibody levels, we found moderate values of correlation, and on average, EIA values were higher than neutralizing antibody values. Others have also reported similar results for the classical PRN assay and the Dade Behring EIA (6). The discrepancies between the enzyme immunoassays and PRMN (PRN) assays may be largely due to the differing formats and MV antigens used in the two tests. Most of the commercial enzyme immunoassays and the Dade Behring EIA kit, in particular, detect antibodies against all viral proteins, while the PRMN (PRN) test detects only functional neutralizing antibodies against specific proteins, i.e., the H and the F proteins. EIAs measure a specific class, for example, only IgG antibodies, while the PRMN (PRN) assay theoretically measures all classes of MV-specific antibodies. Another important issue is the fact that antibodies to the nucleocapsid (N) protein, which do not contribute directly to neutralization, are the most abundant antibodies formed in response to infection and immunization (4). Therefore, it is likely that the EIA predominantly detects antibodies to this antigen, which is also the most abundant protein found in MV-infected cells.
Recombinant viruses expressing various reporter genes, like GFP, luciferase, and β-galactosidase genes, provide us with excellent tools to aid in the assessment of protective neutralizing antibodies (14). For MV, alternatives to the classical PRN assay include H protein-based EIA (3), neutralization EIA, (13) and a neutralization test, based on a recombinant GFP-expressing MV (10). The last study applied a different strain of MV (AIK-C) and a longer and more complicated assay protocol, including a semisolid agarose overlay. This assay did not apply the WHO international anti-MV standards for validation and transformation of the values into mIU/ml, and therefore, it is difficult to compare the two studies or correlate the neutralization values to protection.
In conclusion, this new reporter gene-based PRMN assay provides an expedited and high-throughput alternative to the classical PRN assay. With its delineated protective measures, the validated PRMN assay can be readily transferred to other laboratories and applied to MV immunity testing in larger population-based studies.