In the experimental methods described below, appropriate positive and negative control samples were used to detect any undesirable effects of the procedures.
The index sample was pretreated with RF-Absorbent (Dade Behring, Marburg, Germany), which contains sheep IgM antibodies targeted against human IgG Fc fragments: 250 µl of RF-Absorbent was added to 250 µl of serum, and the mixture was briefly vortexed and incubated for 1 h at room temperature. This procedure, which precipitates IgG along with rheumatoid factor, did not affect the results and excluded insufficient sample pretreatment in the Liaison assays.
Heterophilic antibody interference was excluded by treating the sample, according to the manufacturer's instructions, with a nonspecific antibody-blocking tube (Scantibodies Laboratory, Santee, CA). We also treated the sample by adding 40 µg of PolyMAK-33 (MAK33-IgG1/IgG1 Poly; Roche Diagnostics, Mannheim, Germany) to 250 µl of serum and incubated this mixture for 1 h at room temperature. PolyMAK-33 is a polymerized murine IgG1 preparation, superior to polyclonal mouse immunoglobulins in blocking heterophilic antibody activity (9). This procedure did not affect the results.
However, incubating 250 µl of sample with 75 µl unlabeled beads (kindly provided by DiaSorin) at room temperature for 15 min and centrifuging this mixture for 5 min at 2,000 x g completely eliminated the interference. Apparently, IgM antibodies from the patient reacted with the solid phase in the assays. To test whether this reactivity was restricted to one specific type of microparticle, we evaluated seven different types of microparticles (Dynabeads; Dynal Biotech, Oslo, Norway): M-270 amine, M-270 carboxylic acid, M-270 epoxy, M-280 sheep anti-mouse IgG, M-270 streptavidin, M-280 streptavidin, and M-280 tosyl activated. Two hundred fifty microliters of serum was added to approximately 0.4 x 109 beads, briefly vortexed, and incubated for 15 min at room temperature. After centrifugation (5 min; 2,000 x g), the supernatant was analyzed. The interference could be completely eliminated by using M-280 tosyl-activated beads or M-270 epoxy beads and partially eliminated by using M-270 amine beads, M-280 streptavidin beads, or M-280 sheep anti-mouse IgG beads. No effect was seen when the serum was incubated with M-270 streptavidin beads or with M-270 carboxylic acid beads.
From these experiments, we concluded that (i) since the M-270 streptavidin beads are based on carboxylic acid beads, the described interference would probably not occur in assays applying carboxylic acid-based beads, and (ii) the interference could be completely eliminated with unlabeled tosyl-activated beads but only partially with tosyl-activated beads already covered with an antigen (M-280 streptavidin or M-280 sheep anti-mouse IgG). Increasing the antigen density on the beads might prevent contact of IgM with the bead surface, avoiding the bead-linked interference. However, increasing the antigen density is not easily done; additional chemical blocking could be an alternative solution for the problem. To test this hypothesis, polyvinylpyrrolidone (PVP-360; Sigma-Aldrich) and polyvinyl alcohol (P8136; Sigma-Aldrich) were added to the sample diluents used in the assays (11, 12). This modification indeed strongly reduced or completely eliminated the interference effect. Polyvinylpyrrolidone and polyvinyl alcohol, both water-soluble polymers, can minimize background signals, mainly by competing with nonspecific adsorption of proteins to the solid phase (1, 6, 10).
Since interference was most prominent in the EBV IgM assay, we focused on this assay and found that adding polyvinylpyrrolidone and polyvinyl alcohol to the EBV IgM dilution buffer (buffer A) at final concentrations of 0.1% and 0.005%, respectively, was optimal in reducing the interference. Although this procedure did not completely abolish the strong interference in the index patient, the effect was significantly reduced: the EBV IgM concentration fell from 1,620 mU/liter in the original assay to 320 mU/liter in the modified assay. This easy, inexpensive modification allowed us to estimate the frequency at which this interference occurs by using the modified assay on 120 consecutive samples with a positive result for EBV IgM (>40 mU/liter). The results from this comparative study are shown in Fig. 1 as a Bland-Altman plot. Discrepant results were defined as differing more than 24% (three times the interassay coefficient of variation) between the original and modified assay. Six discrepant results were found (5% of EBV IgM-positive samples; 0.4% of all EBV IgM requests). Interpreting the results as suggested by the manufacturer (<20 mU/liter, negative; 20 to 39 mU/liter, equivocal;
40 mU/liter, positive), one sample became negative and three became equivocal by using this modified assay.
|
Immunoassay interference through endogenous antibodies, such as rheumatoid factor or heterophilic antibodies, remains a continuing challenge (2, 3, 7). The interference we described here was apparently caused by the direct binding of IgM antibodies to surface-modified polystyrene microparticles. These "solid-phase reactive antibodies" have also been described in flow cytometry-based multiplex bead array assays (5, 12). In this technology, the use of reagent blank beads can detect high-level background signals caused by polyreactive antibodies or nonspecific binding antibodies. However, this important advantage is not present in the Liaison assay and many other immunoassay platforms. Since this type of assay interference cannot be predicted or easily recognized in the Liaison assay, the inexpensive preventive measure we propose can reduce the number of false-positive results.