This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowReprints and Permissions
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by van Hoeven, K. H.
Right arrow Articles by Body, B.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by van Hoeven, K. H.
Right arrow Articles by Body, B.

Next Article 

Clinical and Vaccine Immunology, December 2008, p. 1751-1754, Vol. 15, No. 12
1071-412X/08/$08.00+0     doi:10.1128/CVI.00254-08
Copyright © 2008, American Society for Microbiology. All Rights Reserved.

Comparison of Three Enzyme-Linked Immunosorbent Assays for Detection of Immunoglobulin G Antibodies to Tetanus Toxoid with Reference Standards and the Impact on Clinical Practice{triangledown}

Karen H. van Hoeven,1* Connie Dale,1 Phil Foster,2 and Barbara Body2

The Binding Site, Inc., San Diego, California,1 LabCorp, Burlington, North Carolina2

Received 14 July 2008/ Returned for modification 7 August 2008/ Accepted 26 September 2008


arrow
ABSTRACT
 
Accurate determination of the concentrations of immunoglobulin G (IgG) antibody to tetanus toxoid is important in order to evaluate the immunogenicity of tetanus toxoid vaccines, determine immune competence in individual patients, and measure the prevalence of immunity in populations. The performance of three commercially available enzyme-linked immunosorbent assays (ELISAs) for IgG antibodies to tetanus toxoid were evaluated. Serially diluted NIBSC 76/589 and TE-3 human tetanus IgG immunoglobulin international reference standards were analyzed in quadruplicate using ELISAs manufactured by The Binding Site, Inc. (VaccZyme); Scimedx; and Euroimmun. In addition, IgG antibodies to tetanus toxoid were measured in 83 deidentified serum specimens using each manufacturer's ELISA. Each ELISA provided linear results when evaluated with the reference preparations. The Binding Site ELISA provided results that closely corresponded to the reference preparations (y = 1.09x – 0.08), whereas the Scimedx ELISA gave results that were consistently lower (y = 0.21x 0.07) and the Euroimmun ELISA gave results that were consistently higher (y = 1.5x + 0.30) than the reference preparation concentrations. Using the recommended cutoff for each ELISA (<0.10 IU/ml), the overall agreement of all of the ELISA methods was 78%. Three of eighty-three (3.6%) human serum samples demonstrated inadequate immunity with all three assays. The Binding Site ELISA yielded nonprotective antibody concentrations in only these 3 samples, whereas 19 samples (22.9%) according to the Scimedx ELISA and 6 samples (7.2%) according to the Euroimmun ELISA demonstrated nonprotective concentrations. The performance characteristics of ELISAs for tetanus immunoglobulin titers were manufacturer dependent, and the differences translated into important disparities in reported results.


arrow
INTRODUCTION
 
Accurate determination of tetanus toxoid immunoglobulin G (IgG) concentrations is clinically important for evaluating the immunogenicity of tetanus toxoid vaccines (6); determining the immune competence to tetanus in individual patients (5, 8), as part of an evaluation of humoral immune function in general (2); and measuring the prevalence of immunity to tetanus in populations (1, 11).

The gold standard assay for the determination of specific IgG antibodies to tetanus toxoid is the in vivo toxin neutralization test, which is time-consuming, is relatively expensive, is subjective, and raises ethical issues regarding the use of live mammals. The use of accurate and automated in vitro assays is therefore desirable for ethical, clinical, and economic reasons. Moreover, highly reproducible, sensitive, and specific in vitro testing improves the efficiency of the clinical laboratory.

The accurate calibration of these in vitro assays to an internationally recognized reference material is essential for maintaining reproducible and accurate results. The World Health Organization First International Standard for human tetanus immunoglobulin, coded TE-3, was established in 1992, was developed from a pool of 10,628 human plasma donations from Germany, and was calibrated by an international collaborative group from 15 countries representing 15 laboratories (9). Its potency was based on the results of an in vivo toxin neutralization assay in mice that used as its endpoint either death or paralysis (10). The National Institute for Biologic Standards and Control (NIBSC; Hertfordshire, United Kingdom) distributes another reference standard, coded 76/589, consisting of lyophilized pooled human serum and, when it was developed, this standard was also validated against an in vivo toxin neutralization assay.

For the present study, three commercially available enzyme-linked immunosorbent assays (ELISAs) for the measurement of IgG immunoglobulins to tetanus toxoid/toxin were compared by using serial dilutions of these two international standards. In addition, deidentified serum samples were measured with each manufacturer's ELISA, and the results were compared.


arrow
MATERIALS AND METHODS
 
Reference materials. NIBSC reagent 76/589 was supplied by NIBSC (Potters Bar, Hertfordshire, United Kingdom) in a lyophilized vial containing 9.2 IU. It was reconstituted in 9.2 ml of sterile distilled water to yield a working concentration of 1 IU/ml. Serial dilutions of NIBSC 76/589 were performed to yield final concentrations as shown in Table 1.


View this table:
[in this window]
[in a new window]

 
TABLE 1. Dilution protocol for reference standards

An ampoule of the first International Standard for human tetanus immunoglobulin, coded TE-3, was also obtained from the NIBSC. TE-3 was supplied lyophilized at 120 IU, reconstituted in 1 ml of sterile distilled water to yield a working concentration of 120 IU/ml. It was then further diluted to 10 IU/ml by adding 50 µl of the initially reconstituted solution to 550 µl of sterile distilled water. TE-3 was then rediluted to a starting concentration of 7 IU/ml by adding 350 µl of the previously diluted fluid to 150 µl of sterile distilled water. Serial dilutions of TE-3 were performed to yield final concentrations as shown in Table 1. One set of dilutions was made and tested with all three ELISAs.

Serum samples. The ELISAs were compared using 83 serum samples collected in 2007 and submitted for diagnostic testing. These samples would have been discarded but instead were stored at –20°C prior to testing.

Antibody assays. Each of the ELISAs detects IgG antibodies to tetanus toxoid by an indirect technique. Testing was performed on thawed serum samples and reconstituted reference materials in strict accordance with the manufacturers’ specifications using reagents that were supplied with the kits. A DSX 4 Plate ELISA automated processor (The Binding Site, Inc., San Diego, CA) was used in the performance of all assays. Serially diluted NIBSC standards were run in quadruplicate at each dilution. Serum samples were assayed once in accordance with general laboratory procedures. Each run included high and low control solutions provided by the manufacturers.

For Scimedx Corp. (Denville, NJ), the kit lot number was 1300, with an expiration date of January 2008. For VaccZyme by The Binding Site, Ltd. (Birmingham, United Kingdom), the kit lot number was 241679 with an expiration date of November 2008. For Euroimmun (Lübeck, Germany), the kit lot number was E070622AB with a expiration date of 21 June 21 2008. All assays reported in the present study were performed in August 2007, and all laboratory work was performed at LabCorp, Burlington, NC.

Analysis of data. Intra-assay imprecision was calculated as means with a coefficient of variation for the quadruplicate runs. The recovery percentage was calculated by dividing the final result with the expected result and multiplying by 100. Linear regression analysis was performed by using the mean of the quadruplicate results of each reference standard dilution for each ELISA.


arrow
RESULTS
 
Results of controls provided by all manufacturers’ ELISAs were within acceptable limits (Table 2). Intra-assay imprecision ranged from 1.22 to 16.68% for the Scimedx ELISA, from 0.00 to 9.11% for The Binding Site ELISA, and from 2.96 to 24.62% for the Euroimmun ELISA.


View this table:
[in this window]
[in a new window]

 
TABLE 2. Reproducibility of ELISA tests with dilutions of reference standards at levels near the protective antibody titer of 0.1 IU/mla

The Binding Site ELISA provided results that closely corresponded to the reference preparations (1.09x – 0.08), whereas the Scimedx ELISA gave results that were consistently lower (0.2x – 0.07), and the Euroimmun ELISA gave results that were consistently higher (1.5x + 0.30) than the reference preparations (Fig. 1). All assays appeared linear, with correlation coefficients of 0.99 (The Binding Site), 0.97 (Scimedx), and 0.98 (Euroimmun). Because the cutoff for adequate immunity is 0.1 IU/ml for each manufacturer's assay, the results from each of the quadruplicate runs and imprecision at reference values near these clinically important thresholds and at the lower limits of detection are shown in Table 3.


Figure 1
View larger version (14K):
[in this window]
[in a new window]

 
FIG. 1. (A) Concentration of IgG antibodies to tetanus toxoid as detected by reagents from The Binding Site ({blacksquare}), Euroimmun ({blacktriangleup}), and Scimedx ({diamondsuit}) versus serially diluted NIBSC and TE-3 standards. Each point represents the mean of quadruplicate runs. (B) Detail of results obtained at a standard of ≤1.0 IU/ml.


View this table:
[in this window]
[in a new window]

 
TABLE 3. Results of 21 samples showing nonprotective antibody levels with any assay and the corresponding values from all three assaysa

Using the recommended cutoff for each ELISA (<0.10 IU/ml), the overall concordance was 78%, with 62 samples demonstrating adequate anti-tetanus IgG immunoglobulin concentrations. Only 3 of 83 sera demonstrated insufficient immunity with all three assays (Table 3). The Binding Site ELISA yielded insufficient protection in only these 3 samples (3.6%), whereas totals of 19 samples (22.9%) according to Scimedx and 6 samples (7.2%) according to Euroimmun demonstrated inadequately protective concentrations of IgG antibodies.


arrow
DISCUSSION
 
The performance characteristics of the three different ELISA kits were manufacturer dependent. The differences between the assays were manifested as discrepancies in performance against the reference standards and also as variability in the final laboratory results for a number of patient sera that were tested as part of the present study.

The results from the Scimedx ELISA were consistently lower than the reference standards, and a relatively high percentage (23%, or 19 of 83) of the tested serum samples yielded nonprotective antibody titers. The Binding Site ELISA correlated most closely with the reference standards and produced the lowest percentage of nonprotective samples (4% [3 of 83]). The Euroimmun ELISA results were consistently higher than the reference standards, and yet nonprotective antibody titers were demonstrated in 7% of the samples (6 of 83). The explanation of this unexpected paradox is unclear but may be related to the relative avidity of the detected antibodies or the presence of an interfering substance in the samples. Other potential reasons for differences between assays may include the accuracy of the assay calibration and the optimization of the assay protocol on the automated ELISA processor.

The NIBSC distributes TE-3, the World Health Organization International Reference Standard that was derived from German human donor plasma and developed in 15 countries. The three ELISAs tested here were manufactured in unique locations (United States, United Kingdom, and Germany), and the patient samples were from the United States (9). Therefore, no geographical bias should affect the results. No other international reference standard for human tetanus immunoglobulin is currently available to our knowledge, nor are we aware of any equivalent human standard that was subject to such rigorous and stringent development criteria.

The World Health Organization has recommended 0.1 IU/ml as the protective level of specific tetanus toxoid IgG with ELISA testing, and the majority of manufacturers (including the three whose products were evaluated here) have adopted this recommendation (3). Accuracy of testing near the 0.1-IU/ml level is critical because lower concentrations are interpreted as nonprotective, and vaccination may be recommended.

Although testing for tetanus antitoxin levels is frequently used for academic purposes or to test the immunogenicity of vaccines that are in commercial development, specific recommendations have been issued for their appropriate uses in clinical practice. In the 2006 Recommendations of the Advisory Committee on Immunization Practices, there are several guidelines that describe when testing for serum tetanus antitoxin levels would be warranted. For individuals with a history of a prior Arthus reaction after tetanus vaccination, antitoxin levels can be obtained if the last vaccine was administered at least 10 years earlier (4, 7). Among adults or adolescents who probably received prior vaccination but for whom records are not available, serologic testing may also be considered to avoid unnecessary vaccination (4, 7). In the 2005 Practice Parameters for the Diagnosis and Management of Primary Immunodeficiency, it was stated that the determination of antibody levels to tetanus vaccination are often performed as one component of an evaluation of a specific immune response (2).

In the assessment of immune competence, pre- and postvaccination antibody levels may be compared, and a fourfold increase is generally considered desirable. Accuracy as well as reproducibility at the lower levels is therefore critical when immunodeficient patients are monitored because serial samples may be drawn over time to assess the efficacy of therapeutic immune globulin treatment and the potential development of immune competence.

The present study provides valuable comparative data for laboratories that are evaluating different manufacturers’ products for routine use in their own centers. Each product that was evaluated here performed well relative to internal precision, linearity, and internal controls. However, significant differences were shown relative to international reference preparations. Further work may be needed to confirm these findings, particularly if manufacturing processes or standards evolve or change over time.


arrow
ACKNOWLEDGMENTS
 
We gratefully acknowledge the donation of the ELISA reagents by Euroimmun and The Binding Site for the testing that was performed in this study.

K.H.V.H. and C.D. are employees of The Binding Site, Inc., which produces one of the assays evaluated in this study.


arrow
FOOTNOTES
 
* Corresponding author. Mailing address: The Binding Site, Inc., 5889 Oberlin Dr., Ste. 101, San Diego, CA 92121. Phone: (267) 402-7569. Fax: (413) 612-9683. E-mail: karen.vanhoeven{at}thebindingsite.com Back

{triangledown} Published ahead of print on 8 October 2008. Back


arrow
REFERENCES
 
    1
  1. Alagappan, K., W. Rennie, D. Lin, and C. Auerbach. 1998. Immunologic response to tetanus toxoid in the elderly: one-year follow-up. Ann. Emerg. Med. 32:155-160.[CrossRef][Medline]
  2. 2
  3. Bonilla, F. A., I. L. Bernstein, D. A. Khan, Z. K. Ballas, J. Chinen, M. M. Frank, L. J. Kobrynski, A. I. Levinson, B. Mazer, R. P. Nelson, Jr., J. S. Orange, J. M. Routes, W. T. Shearer, and R. U. Sorensen. 2005. Practice parameter for the diagnosis and management of primary immunodeficiency. Ann. Allergy Asthma Immunol. 94:S1-S63.[Medline]
  4. 3
  5. Borrow, P., P. Balmer, and M. H. Roper. 2007. The immunologic basis for immunization series, module 3. Tetanus update 2006. WHO Press, Geneva, Switzerland. http://whqlibdoc.who.int/publications/2007/9789241595551_eng.pdf.
  6. 4
  7. Broder, K. R., M. M. Cortese, J. K. Iskander, K. Kretsinger, B. A. Slade, K. H. Brown, C. M. Mijalski, T. Tiwari, E. J. Weston, A. C. Cohn, P. U. Srivastava, J. S. Moran, B. Schwartz, and T. V. Murphy. 2006. Preventing tetanus, diphtheria, and pertussis among adolescents: use of tetanus toxoid, reduced diphtheria toxoid and acellular pertussis vaccines. MMWR Recomm. Rep. 55(RR03):1-34.[Medline]
  8. 5
  9. de La Chapelle, A., O. Lavabre, M. Pinsard, J. Delamonica, and E. H. Relyveld. 2002. Tetanus in a renal transplant recipient exhibiting the presence of circulating antitetanus antibodies determined by ELISA. Biomed. Pharmacother. 56:208-210.[CrossRef][Medline]
  10. 6
  11. Grimprel, E., F. von Sonnenburg, R. Sanger, V. Abitbol, J. M. Wolter, and L. M. Schuerman. 2005. Combined reduced-antigen-content diphtheria- tetanus-acellular pertussis and polio vaccine (dTpa-IPV) for booster vaccination of adults. Vaccine 23:3657-3667.[CrossRef][Medline]
  12. 7
  13. Kretsinger, K., K. R. Broder, M. M. Cortese, M. P. Joyce, I. Ortega-Sanchez, G. M. Lee, T. Tiwari, A. C. Cohn, B. A. Slade, J. K. Iskander, C. M. Mijalski, K. H. Brown, and T. V. Murphy. 2006. Preventing tetanus, diphtheria, and pertussis among adults: use of tetanus toxoid, reduced diphtheria toxoid and acellular pertussis vaccine. MMWR Recomm. Rep. 55(RR17):1-37.[Medline]
  14. 8
  15. Mayorga, C., M. J. Torres, J. L. Corzo, J. Alvarez, J. A. Garcia, C. A. Rodriguez, M. Blanca, and A. Jurado. 2003. Immediate allergy to tetanus toxoid vaccine: determination of immunoglobulin E and immunoglobulin G antibodies to allergenic proteins. Ann. Allergy Asthma Immunol. 90:238-243.[Medline]
  16. 9
  17. National Institute for Biologic Standards and Control. 2008. WHO International Standard. First International Standard for Tetanus Immunoglobulin, human NIBSC code: TE-3. NIBSC, Potters Bar, Hertfordshire, United Kingdom. http://www.nibsc.ac.uk/documents/ifu/TE-3.pdf.
  18. 10
  19. Sesardic, D., M. Y. Wong, R. E. Gaines Das, and M. J. Corbel. 1993. The first international standard for antitetanus immunoglobulin, human: pharmaceutical evaluation and international collaborative study. Biologicals 21:67-75.[CrossRef][Medline]
  20. 11
  21. Shohat, T., E. Marva, Y. Sivan, I. Lerman, A. Mates, and A. Cohen. 2000. Immunologic response to a single dose of tetanus toxoid in older people. J. Am. Geriatr. Soc. 48:949-951.[Medline]


Clinical and Vaccine Immunology, December 2008, p. 1751-1754, Vol. 15, No. 12
1071-412X/08/$08.00+0     doi:10.1128/CVI.00254-08
Copyright © 2008, American Society for Microbiology. All Rights Reserved.





This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowReprints and Permissions
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by van Hoeven, K. H.
Right arrow Articles by Body, B.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by van Hoeven, K. H.
Right arrow Articles by Body, B.