Clinical and Vaccine Immunology, September 2007, p. 1094-1101, Vol. 14, No. 9
1071-412X/07/$08.00+0 doi:10.1128/CVI.00183-07
Copyright © 2007, American Society for Microbiology. All Rights Reserved.

Departments of Medicine, Microbiology, and Immunology, Albert Einstein College of Medicine, Bronx, New York,1 Fordham University, Bronx, New York2
Received 30 April 2007/ Returned for modification 21 June 2007/ Accepted 3 July 2007
|
|
|---|
|
|
|---|
SEB is a well-characterized 28-kDa protein that is most closely related to SEC and the streptococcal pyrogenic exotoxins A and C (12, 33). Like all of the aforementioned toxins, SEB is a superantigen and is one of the most potent mitogens described. SEB mediates its biological effects by binding to the major histocompatibility complex (MHC) class II complex at a distinct site and is different from other antigens in that it does not have to be preprocessed. The toxin is presented to a T-cell antigen receptor by an MHC class II molecule, forming ternary complexes that trigger cytokine production and T-cell proliferation (5, 13, 18, 33).
During the 1960s, when the United States had an offensive biological warfare program, SEB, then code named PG, was studied extensively as a biological incapacitant. The toxin was especially attractive as a biological agent because much lower quantities of SEB than of synthetic chemicals were needed to produce intoxicating effects. The dose of SEB that is incapacitating for 50% of the human population exposed to SEB was predicted to be 0.0004 µg/kg of body weight, and the 50% lethal dose was predicted to be 0.02 µg/kg by both the inhalational and the intravenous routes (36). SEB can easily be synthesized in large quantities and is considered a major agent of biological warfare. SEB is currently listed as a category B select agent (36).
Staphylococcus sepsis is very common and is associated with high rates of mortality and high costs in affected patients (29). The presence of genes encoding for SEB and several other bacterial superantigens in clinical S. aureus isolates has been described (10, 11). However, the contribution of SEB production to the outcome of staphylococcus-related diseases is not known. This is of particular importance, because the effects of toxins could potentially be neutralized with specific immune globulin therapy (2, 14, 15). Only a few clinical studies have actually demonstrated the presence of SEB in the serum of patients with staphylococcal infections (1), presumably because of the lack of highly sensitive detection methods. Most commercially available diagnostic tests detect SEB in the nanogram range (20, 22, 30-32, 34) Although mass spectrometry has created a niche for the analysis of proteinaceous toxins, its main drawbacks are its sophisticated instrumentation and its high costs (16). Hence, sensitive immunoassays continue to provide a more realistic alternative. We report on a highly sensitive capture enzyme-linked immunosorbent assay (ELISA) that can be used to further investigate the pathogenesis and treatment of SEB-induced disease.
|
|
|---|
MAbs. Monoclonal antibodies (MAbs) to SEB were generated in the Hybridoma Facility of the Cancer Center at the Albert Einstein College of Medicine. Three of the MAbs, 10F1 (immunoglobulin A [IgA]), 17C12 (IgG2a), and 20B1 (IgG1), were further used in these studies. A concentrated hybridoma supernatant containing MAb was used as a source of antibody. The antibody concentrations were determined by ELISA by the use of commercial isotype standards (MP Biomedicals) of known concentration.
Staphylococcus aureus strains. S. aureus isolates from bacteremic patients were obtained from the microbiology laboratory at the Montefiore Medical Center (Bronx, NY). Isolates were maintained on blood agar plates and grown in brain heart infusion (BHI) medium (Bacto-BD) at 37°C with shaking. DNA was isolated with a commercially available kit (QIAGEN Inc.), following the manufacturer's instructions. The presence of the SEB gene in clinical strains was determined by PCR with previously published primers and by previously published protocols (9). Genetically engineered S. aureus strains (strains RN6734, PRN7114/RN9432, and PRN7116/RN9469) were kindly provided by R. P. Novick (Skirball Institute, New York University) and were used as control strains. These strains are described in detail elsewhere (37). RN6734 is an enterotoxin-negative strain, and PRN7114 contains an intact SEB gene on a plasmid, while PRN7116 was transformed with a plasmid containing a truncated SEB gene and, consequently, does not produce SEB. In addition, clinical S. aureus strains from our own collection were used; these included strains 2, 9, and 36, which are methicillin-resistant and SEB-positive strains, and strain 18, which is a methicillin-resistant SEB-negative strain.
Capture ELISAs. The ELISAs were performed in 96-well polystyrene plates (Corning, Corning, NY). Capture ELISAs were configured in a double-sandwich format, similar to that used in previously published assays (4, 35), in which the "bottom" MAb was used to capture the antigen (SEB) and the "top" MAb was used to detect SEB. We developed two separate assays, both with the same "bottom" antibody but each with a different "top" or "detection" antibody of different isotypes. The protocols are as follows. First, the polystyrene plates were coated for 1 h at room temperature (RT) with 100 µl of 1 µg/ml goat anti-mouse IgA polyclonal antibody diluted in 0.02 M PBS (pH 7.2). Second, the plates were blocked with 5% milk in PBS overnight at 4°C. Third, the "bottom" or "capture" antibody, MAb 10F1 (IgA), was added at a volume of 100 µl per well at a concentration of 1 µg/ml diluted in 1% bovine serum albumin (BSA) in PBS (1% BSA-PBS). The plate was incubated for 30 min at RT and was then washed three times in an automated microtiter plate washer (Skanwasher 400; Molecular Devices). All washes were performed with PBS containing 0.1% Tween 20 and were interrupted by 1-min soaks. Fourth, the samples were added to each well and serially diluted across 12 wells. Purified SEB (1 µg/ml) diluted in 1% BSA-PBS was used as the concentration standard. The samples and the SEB standard were incubated for 1 h at RT. Fifth, the "top" or "detection" MAb, MAb 20B1 (IgG1) or MAb 17C12 (IgG2a), was added at 1 µg/ml in 1% BSA-PBS, and the plates were incubated for 30 min at RT. Sixth, 1 µg/ml of alkaline phosphatase-conjugated goat anti-mouse IgG1 or IgG2a in 1% BSA-PBS for MAb 20B1 or 17C12, respectively, was added for 30 min at RT. Lastly, for colorimetric detection, 50 µl per well of phosphatase substrate (1 mg/ml p-nitrophenylphosphate in 0.001 M MgCl2 and 0.05 M Na2CO3, pH 9.8) was added for 15 to 30 min at RT. The plates were read at 405 nm with a microtiter plate reader (Emax micro plate reader; Molecular Devices).
ELISA specificity for SEB. To determine the specificity of the ELISA for SEB compared with its specificity for other staphylococcal enterotoxins, we diluted purified SEB, SEC1, SEA, and TSST-1 to a concentration of 2 µg/ml in assay diluent (BSA-PBS) and measured these samples in the ELISA described above.
Staphylococcal protein A interference. Purified SEB was diluted to 1 µg/ml into a solution of recombinant staphylococcal protein A (Pierce) at concentrations ranging from 1,000 to 15.6 ng/ml in assay diluent (BSA-PBS). The samples were assayed by the capture ELISA with the anti-SEB specific antibodies described above, and the values were compared to those for SEB diluted in assay diluent alone to determine possible interference from protein A. Recombinant protein A was used to avoid possible enterotoxin contamination.
SEB detection in culture supernatants and whole-cell lysates. Single colonies from the S. aureus isolates were grown in BHI medium overnight. For SEB quantification, these cultures were diluted into fresh medium to an A600 of 0.1 and were cultured for an additional 24 h. Over the first 7 h, aliquots of the supernatant were collected hourly and the absorbance at 600 nm was measured. The time period of supernatant sampling corresponded to the logarithmic growth phase of the cultures. All supernatant samples were frozen, and the SEB concentration was determined by the capture ELISA. For the detection of SEB in whole-cell lysates, approximately 10 µl of cells was scraped directly from a blood agar plate. The cells were boiled in 100 µl of PBS for 30 min and then cooled to 4°C. The cellular debris was pelleted by centrifugation, and the SEB concentration in the resulting supernatant was measured by ELISA.
Analysis of biological specimens. The ability to accurately measure SEB in biological fluids was determined by diluting a known amount (1 µg/ml) of SEB in solutions of 50% human serum, 50% mouse serum, or 15 mg/ml mouse feces. The levels in these samples, as well as those in a standard sample of SEB diluted in assay diluent, were measured by the ELISA described above. Prior to these experiments, the biological fluids used were analyzed for the presence of preexisting anti-SEB antibodies by a direct-binding ELISA and were determined to have nondetectable levels (data not shown).
Mouse experiments. BALB/c mice (NCI) (n = 5 to 10 mice per group) were challenged by intraperitoneal (i.p.) injection, oral gavage feeding, or intranasal (i.n.) administration of SEB. Mice were injected i.p. with either 25 mg of D-galactosamine or 70 µg of lipopolysaccharide (LPS; Sigma) in PBS, followed by 50 µg of purified SEB dissolved in 100 µl of PBS. HLA-DR3 transgenic mice (27) (a generous gift of C. S. David at the Mayo Clinic) were also injected i.p. with 50 µg of purified SEB dissolved in 100 µl of PBS. This concentration was suggested by other investigators to be the 100% lethal dose within 48 h for mice (25). Sera were collected by retro-orbital bleeding at 8 h postinjection and frozen. Urine release was induced by rubbing the mouse's lower abdomen, and the urine was collected in a microcentrifuge tube. The mice were also fed 200 µg of SEB in PBS by oral gavage. The mice developed mild symptoms of intoxication (decreased mobility) within 24 h but survived. For i.n. administration, the mice were briefly anesthetized and 20 µg of SEB in 20 µl of PBS was placed on their nostrils, which allowed inhalation of the toxin. These mice developed no symptoms. SEB levels were also measured in the serum of mice 2, 8, and 24 h after infection with 109 CFU of clinical S. aureus strains that were SEB positive (strain 36) and SEB negative (strain 18).
Statistical analysis. Student t tests, standard curves, concentrations, and trend lines were generated and calculated by using Excel X software (Microsoft Corporation).
|
|
|---|
|
View this table: [in a new window] |
TABLE 1. Sensitivities of different capture ELISAs tested
|
![]() View larger version (19K): [in a new window] |
FIG. 1. Configuration of capture ELISA for SEB detection. GAM, goat anti-mouse; AP, alkaline phosphatase; UNLB, unlabeled.
|
![]() View larger version (11K): [in a new window] |
FIG. 2. SEB detection by two different MAb sandwich ELISAs. (a) MAbs 10F1 and 20B1; (b) MAbs 10F1 and 17C12. A comparison of SEB detection in 50% murine serum, 50% human serum, feces, and BSA is shown. The test was performed in triplicate, and error bars represent standard deviations.
|
|
View this table: [in a new window] |
TABLE 2. Results of ELISAs with MAbs 10F1 and 20B1 and MAbs 10F1 and 17C12
|
![]() View larger version (44K): [in a new window] |
FIG. 3. (a) Measurement of equal amounts of SEA, SEC, TSST-1, and SEB by ELISA (with MAbs 10F1 and 20B1). As expected, some cross-reactivity is observed at the highest dose of SEC, which is closely related to SEB. (b) Measurement of SEB by ELISA in the presence of increasing staphylococcal protein A concentrations. The test was performed in triplicate, and error bars represent standard deviations. Note that the measurements were affected only by the presence of high levels of protein A (1,000 ng/ml), as the capture ELISA measures slightly increased SEB levels compared to the levels measured for the control.
|
![]() View larger version (20K): [in a new window] |
FIG. 4. SEB was measured by capture ELISA in diluted supernatants of three SEB-positive S. aureus strains (strains 2, 9, and PRN7114) after overnight growth in BHI medium at 37°C. The supernatants from S. aureus strains that contained either no SEB gene or a defective gene were used as controls (PRN7116/RN9469, RN6734). Beginning with undiluted samples, the supernatant was serially diluted 1:3 across 12 wells. SEB-negative strains demonstrated background detection (A405, approximately 0.06) for this ELISA. Note that more SEB is produced by PRN7114.
|
(ii) SEB detection in body fluids of SEB-exposed and SEB-positive S. aureus-infected mice. The capture ELISA was used for the detection of toxin in body fluids. First, mice were injected i.p. with a lethal dose of 20 µg purified SEB after sensitization with D-galactosamine. Serum and urine were collected 8 h after injection. Serum SEB levels were variable and ranged from 28 to 150 ng/ml (median, 74 ng/ml). The SEB levels in the urine 8 h after i.p. injection ranged from 10 to 47 ng/ml (median, 40 ng/ml). Presumed transcytosis of SEB across mucosal membranes was measured in models of airborne and food-borne SEB exposure (Fig. 5a). SEB (200 µg) was fed to the mice via a gavage needle. In a different experiment, 20 µg of purified SEB was given i.n. (Fig. 5b). SEB levels were measured by ELISA in the stool and blood at 2, 16, 24, and 36 h and in the blood after i.n. exposure at 2, 8, 24, and 36 h. SEB was consistently detected in the stools of mice fed SEB. Transcytosis of SEB from the gut, as well as from the alveolar space, into the bloodstream was rapid and could be documented as early as 2 h postexposure. As expected, the serum SEB levels after i.n. challenge were lower than those measured in serum after oral challenge with a 10-fold higher SEB dose. Next, we infected mice intravenously with 109 CFU from SEB-producing and non-SEB-producing (no SEB gene) S. aureus clinical strains. SEB was detected in three of five mice that were infected with an SEB-positive strain and in none of the mice that were infected with the SEB-negative strain. The serum SEB levels were higher at 2 h postinfection (10.3 ± 7.4 µg/ml) than at 24 h postinfection (2.0 ± 1.3 µg/ml), consistent with the higher average numbers of CFU (log 4.5 at 2 h versus log 3.9 at 24 h) early in the course of the infection.
![]() View larger version (14K): [in a new window] |
FIG. 5. Measurement of SEB by ELISA in body fluids of challenged mice. The average SEB levels that were measured by capture ELISA in the sera and stools of BALB/c mice after oral and i.n. challenges with purified SEB are shown. Error bars reflect standard deviations (n = 5 to 10 mice per group). (a) SEB levels detected in the sera and stools after oral challenge with 200 µg SEB at the given times; (b) SEB can also be detected in the serum of mice that were challenged with a 10-fold lower dose (20 µg) of SEB i.n. Prior to intoxication, SEB was not detectable in the sera or the stools of the mice (data not shown).
|
![]() View larger version (22K): [in a new window] |
FIG. 6. Comparison of serum SEB levels in different mouse models for SEBILS. Shown are the average SEB levels in the sera (a) and urine (b) of BALB/c mice, as measured by capture ELISA, after i.p. injection of (i) 50 µg SEB only, (ii) SEB + 25 mg D-galactosamine, or (iii) SEB plus 75 µg LPS and (iv) average levels in serum of transgenic HLA-DR3 mice after i.p. challenge with 50 µg SEB. Error bars denotes standard deviations within a group of mice (n = 5 to 10 mice). *, significant differences in SEB levels by Student t test (P < 0.05). Note that the peak SEB levels in serum differ significantly between HLA-DR3 and BALB/c mice, as SEB appears later in the serum of HLA-DR3 mice (a). At 8 h SEB levels in urine are also significantly higher in HLA-DR3 mice than in BALB/c mice (b).
|
|
|
|---|
In general, SEB toxin levels are expected to be very low, in the range of pg/ml, in the serum of exposed humans who either survived or are symptomatic from SEB intoxication (36). The capture ELISA described here is specific and highly sensitive for the detection of toxin in the pg/ml range. Neither culture supernatants nor serum, urine, or stool products interfered with the performance of this assay. The extent to which preexisting antibodies in the serum of already exposed patients could interfere with the sensitivity of the ELISA is unknown at this point. Our data indicate that SEB-injected mice develop significant levels of SEB in their serum after 8 h and excrete the toxin renally within that time frame. In addition, our results demonstrate that transcytosis of ingested or inhaled SEB occurs across mucosal barriers within a short time. In contrast to other studies, we were able to detect the presumed transcytosis of SEB into serum by ELISA, whereas others demonstrated transcytosis only by biological assays, which would not be useful in the clinical setting (8). Given that we challenged mice with the same dose used in other published studies, we believe that our ability to detect SEB was most likely a result of the enhanced sensitivity of our test.
Up to 30% of the human population is colonized with S. aureus, and SEB is produced by 5 to 10% of clinical S. aureus strains (6, 11). Humans colonized with toxin-producing S. aureus often exhibit neutralizing antibodies (11). The protective efficacy of human antibodies in the situation of a massive exposure following an attack with SEB toxin is not known. Although neutralizing IgM antibodies may be available in preimmune subjects, intoxication with a superantigen in the setting of a high level of SEB exposure could potentially interfere with the mobilization of memory B cells. This ELISA will be useful in studying these aspects as well as studying the effects of reagents that protect transcytosis across mucosal barriers.
SEB is thought to rapidly bind to the MHC and T-cell antigen receptor in the host, and hence, the design of an effective therapeutic intervention remains a challenge. SEB has a much higher affinity for human MHC than mouse MHC, and consequently, murine animal models have their limitations in answering these important questions. To enhance the toxicity of SEB in murine models, D-galactosamine, an agent toxic for the liver, is usually used (18). The exact mechanism by which this chemical affects SEB levels is not known, but it has been suggested that it may augment liver toxicity and cytokine levels (19). More recent studies have proposed the use of mice that are transgenic for HLA-DR3, which renders them more sensitive to SEB intoxication (27, 28). Our data indicate that the serum SEB levels in these murine models vary over a considerable range and that the clearance of SEB from the serum of D-galactosamine- and LPS-cotreated mice may be significantly different from that from the serum of HLA-DR3 transgenic mice. More extensive studies will have to be done to determine which model best mimics the clearance of the toxin in humans. This is of great importance in settings in which drugs that prevent the rapid binding of this toxin to MHC and T-cell receptors are to be developed (24, 25, 27, 28).
S. aureus is the second most common pathogen recovered in nosocomial bloodstream infections in the United States (38). The regulation of staphylococcal enterotoxins has been studied predominately in in vitro experiments, in which toxin levels are determined by protein precipitation of the supernatant (37). Recent studies reveal that the in vivo regulation may not correlate with the in vitro regulation of the toxin (21, 23), possibly because the pathogen burden and/or the in vivo growth conditions potentially affect toxin production (39-41). The contribution of SEB to the outcome of staphylococcal infection is still unclear, mainly because tools for the measurement of low SEB toxin levels in serum and tissue are lacking (1, 6). Even if only a subset of S. aureus infections are caused by SEB toxin-producing strains, these may be more likely to cause lethal disease and may be more likely to infect patients who do not have protective antibodies. Accordingly, although several studies have shown that vaccination can prevent toxin-mediated disease, no systemic clinical investigations with patients have been undertaken (3). Our ELISA is more sensitive than other published SEB detection assays and may be helpful in future clinical investigations.
This work was funded by Northeast Biodefense Center (grant U54-AI057158 and grant 5U54AI057158-05 to P. I. Lipkin). Part of this work was done in collaboration with the Northeast Biodefense Center Monoclonal Antibody Core (P. I. Scharff).
Published ahead of print on 18 July 2007. ![]()
|
|
|---|