Clinical Value of Multiplexed Bead-Based Immunoassays for Detection of Autoantibodies to Nuclear Antigens
Erik Avaniss-Aghajani,*
Sophia Berzon, and
Arlen Sarkissian
Primex Clinical Laboratories, Inc., 16742 Stagg St. #120, Van Nuys, California 91406
Received 15 January 2007/
Returned for modification 16 February 2007/
Accepted 9 March 2007

ABSTRACT
The advent of multiplexed bead assays in recent years has introduced
a new dimension of testing for complex diseases such as lupus,
which can involve multiple autoantibodies. The ability to rapidly
identify multiple autoantibodies, with high sensitivity and
specificity in an automated fashion, is highly attractive. The
aim of this study was to assess the performance and clinical
value of multiplexed bead-based (AtheNA Multi-Lyte ANA-II test
system) immunoassays both by comparing the results with those
achieved by indirect fluorescent-antibody assay (IFA) or conventional
enzyme immunoassays (EIAs) and by independent identification
of autoantibodies in well-characterized samples. To achieve
this goal, 984 samples were tested for seven analytes (SS/A,
SS/B, Sm, RNP, Scl-70, double-stranded DNA [dsDNA], and centromere
B) in both traditional and bead-based assays. The average concordance
for the different analytes was 91%, ranging from 81% (dsDNA)
to 97% (centromere B). The average relative specificity and
sensitivity for the analytes were also high, 92% and 81%, respectively.
An examination of 93 "normal controls" demonstrated a 7% false-positive
rate, which was comparable to IFA. Percentages of different
autoantibodies found in patients with a variety of disease conditions
(34 with calcinosis, Raynaud's phenomenon, esophageal dysmotility,
sclerodactyly, and telangiectasia; 41 with mixed connective
tissue disease; 24 with scleroderma; and 35 with Sjogren's syndrome)
were well within the range expected from each group. A scrutiny
of results from AtheNA and EIA and Farr results for 185 systemic
lupus erythematosus samples revealed comparable results by both
methods, with the exception of SS/A and dsDNA, where AtheNA
had a higher percentage of SS/A-positive results compared to
EIA (51% versus 29%) and a lower percentage of dsDNA-positive
results (18% versus 28% at a cutoff of 5 IU/ml).

INTRODUCTION
Detection of antinuclear antibodies (ANAs) has a significant
role in diagnosis and prognosis for clinically indicated patients
with a variety of autoimmune vascular diseases. Traditionally,
the "gold standard" test for detection of ANAs has been the
indirect fluorescent-antibody assay (IFA). The advent of using
the human HEp-2 cell line for detection of ANAs in the past
20 years has provided sensitivity and brought more standardization
and therefore acceptance of this test globally (
11). It provided
superior resolution for detection of different staining patterns
that was not available before. The increased sensitivity also
brought forth a more reliable use of titer cutoffs for determining
positive results. Though reliable, ANA testing by IFA has had
its share of problems and criticism over the years. The test
has been deemed "subjective" and highly dependent on the competence
of the technician reading the slides (
10). IFA testing is also
an issue for high-volume laboratories performing ANA screens
routinely. To circumvent these problems, researchers have evaluated
ANA screening methods using enzyme immunoassays (EIAs) that
are usually prepared from HEp-2 cells. A few studies have shown
comparable sensitivity and specificity with IFA testing (
6,
7). The advocates of EIA-ANA testing herald the objectivity
of the results and the ability to automate and run multiple
samples reliably (
1). Most critics, though, cite issues with
sensitivity (either too sensitive, resulting in a high number
of false positives, or the opposite, resulting in false negatives)
(
2), the range of specificity (number of different extractable
nuclear antigens [ENAs] detected), and the lack of the ability
to detect different patterns available by IFA. There have been
suggestions that these shortcomings can be overcome by testing
each sample with EIAs specific for testing single ENAs. This
idea, although theoretically sound, defeats the purpose of ANA
testing by EIA, since it considerably increases cost and time
to result, not to mention erroneous diagnosis based on a single
test result.
Advances in technology have recently provided a new methodology for ANA testing (8, 12, 14). Fluorescent bead-based flow cytometry, pioneered by Luminex Inc., has allowed different manufacturers to produce kits capable of detecting multiple autoantibodies to ENAs simultaneously. Recent papers have evaluated these kits against IFA and EIA and shown excellent concordance, sensitivity, and specificity (12, 14). We compared the results of 984 clinical samples that were originally tested by IFA for ANA screen by EIA (Inova) for anti-Ro (SS-A), anti-La (SS-B), anti-Sm, anti-RNP, anticentromere, anti-Jo1, anti-Scl70, and antihistone and by the Farr method for dsDNA with AtheNA Multi-Lyte, a multiplexed microparticle immunoassay for antibodies to ANAs. Our goals were to determine if these systems could reliably be used as a screen to replace ANA testing by IFA. Furthermore, we sought to determine if the test was specific and sensitive enough to be used for monitoring flares for patients that are monitored by rheumatologists.

MATERIALS AND METHODS
Clinical samples.
A total of 984 clinical samples were obtained from Rheumatology
Diagnostics Laboratory (Los Angeles, CA). These samples were
in three distinct groups. One group consisted of 452 samples
from patients who had well-documented disease states (systemic
lupus erythematosus [SLE]; Sjogren's syndrome; scleroderma;
calcinosis, Raynaud's phenomenon, esophageal dysmotility, sclerodactyly,
and telangiectasia [CREST]; Raynaud's phenomenon; drug-induced
lupus; and mixed connective tissue disease [MCTD]) and test
results by different methodologies. Another group consisted
of 438 samples which had only test results by different methodologies.
The third group consisted of 94 patient samples that were deemed
to be "normal controls" or at the time of their visit with the
physician did not demonstrate any symptoms or clinical signs
of illness. All samples obtained were only identified by a number
to meet the Health Information Portability and Accountability
Act patient confidentiality guidelines.
ANA screen and centromere ENA detection by IFA.
The RhiGene Titer-Fluor ANA test system was used as the ANA screening methodology for the obtained samples. The RhiGene Titer-Fluor ANA test system is an indirect fluorescent-antibody assay utilizing HEp-2 tissue culture cells as a substrate for the qualitative and/or semiquantitative determination of antinuclear antibodies in human serum. A titer calibrator was utilized by preparing eight serial dilutions to incorporate a range of 1:2,560 to <1:40. After processing the HEp-2 slides according to the manufacturer's suggested protocol, the Rhigene ImageTiter workstation was used to determine the appropriate titer for each sample. Samples with titers of 1:40 or greater were considered positives. The same methodology was used for detection of centromere patterns. Similarly, samples with titers of 1:40 and a centromeric pattern were determined to be positive for centromere autoantibodies.
ENA assays.
Inova's QUANTA Lite enzyme-linked immunosorbent assays (ELISAs) were used for the semiquantitative detection of Sm, RNP, SS-A (60 kDa and 52 kDa), SS-B, and Scl-70 antibodies in the serum. Purified Sm-RNP, SS/A, SS/B, and Scl-70 antigens were bound to the wells of a polystyrene microwell plate under conditions that preserve the antigen in its native state. The manufacturer's suggested protocol was used for all of the assays. The following formula was used for calculating the sample results: sample value = (sample optical density/test low positive) x test low positive, where test low positive refers to the low value of the positive control provided with the kit. Samples with values of 20 or above by using the above formula were considered positives.
dsDNA antibodies by the Farr technique.
DPC's anti-DNA radioimmunoassay was used for detection of antibodies against double-stranded or native DNA. The assay is based on principles of the Farr technique (4). The assay utilizes 125I-labeled recombinant DNA for detection of antibodies against double-stranded DNA. The concentration of the DNA-bound antibodies in the serum is determined by readings against a calibration curve ranging from 0 to 50 IU/ml. A 25-µl volume of undiluted serum was used for this assay. The manufacturer's suggested protocol was followed for performing this assay. All samples with results exceeding 5 IU/ml were considered positive samples.
ANA detection by multiplexed bead-based immunoassay.
We chose the AtheNA Multi-Lyte ANA-II test system for testing our 984 samples. AtheNA Multi-Lyte utilizes a suspension that contains separate distinguishable 5.6-µm polystyrene beads that are conjugated with the following autoantigens: SS/A, SS/B, Sm, snRNP B/B', U1 snRNP 68, U1 snRNP A, U1 snRNP C, Scl-70, Jo-1, centromere B, dsDNA, histone H, and histone HLY. Samples were diluted 1:21 and incubated with the bead mixture for 30 min. Samples were then washed to remove all unbound serum components. Phycoerythrin-conjugated goat anti-human immunoglobulin G (y chain specific) was added to the samples, the samples were incubated for another 30 min, and the samples were read on a Luminex 100 system. The manufacturer's suggested protocol was followed for performing this assay. All samples with results exceeding 120 arbitrary units (AU)/ml were considered positive samples. Samples below 100 AU/ml were considered negative. All samples with results between 100 and 120 AU/ml were considered equivocal and were repeated for resolution. Samples that stayed in the equivocal range after repeat testing were excluded from our study.

RESULTS
Concordance, relative specificity, and relative sensitivity of bead-based assay to ELISA and Farr.
A total of 984 serum samples were compared to determine the
concordance of the AtheNA Multi-Lyte assay with the traditional
individual ELISAs for individual ENAs and dsDNA by Farr. All
of the 984 samples were tested by the AtheNA system, but a fraction
of these samples had appropriate test results for each individual
antibody. The number of samples with test results for Jo-1 (
n = 6) and histone (
n = 47) were far below the number of samples
with results for other assays and therefore they were eliminated
from this study. The results of the concordance between the
other seven assays and their relative specificities and sensitivities
based on the agreement numbers are detailed in Table
1. The
results obtained by the AtheNA system compared well with the
results from the individual ELISA tests and the Farr assay.
SS/B, Sm, centromere B, and Scl-70 had a greater than 92% concordance
and specificities in excess of 95%. SS/A and RNP had lower levels
of concordance (86% and 87%, respectively) but still demonstrated
greater than 88% relative specificity. dsDNA was least concordant
(81%). A cutoff value of 5.0 IU/ml was chosen when the samples
were originally run by the Farr assay. The concordance increases
to 83% if the cutoff of 10 IU/ml is used.
View this table:
[in this window]
[in a new window]
|
TABLE 1. Concordance and relative sensitivity and specificity of Wampole's AtheNA Multilyte compared to Inova's Quanta Lite EIA and DPC's anti-DNA radioimmunoassay
|
Determination of false positive rate.
Ninety-four "normal control" samples were screened by both AtheNA
and IFA (Table
2). Seven samples (7.4%) were positive by both
IFA and AtheNA. The other 87 samples were negative with both
AtheNA and IFA. Four of the seven positive samples were just
barely positive by IFA, with titers of 1:40. Two samples had
high titers of 1:640. Further investigation of these high-titer
samples showed high SS/A results by AtheNA. All seven samples
had only one antibody present (either SS/A or RNP) when tested
by both AtheNA and conventional methods. The frequency of these
positive results in "normal control" specimens was well within
the range previously reported in the literature (
11,
15).
Determination of false-negative rate.
A total of 137 well-characterized SLE samples with negative
AtheNA results were scrutinized for determination of the false-negative
rate. Of these samples, 124 were also negative for all autoantibodies
by ELISA and for dsDNA by Farr. Out of the remaining 13, 1 (0.7%)
was positive for SSA, 4 (2.9%) were positive for RNP, and 8
(5.8%) were positive for dsDNA by Farr (cutoff, 10 IU/ml).
Determination of individual ENA frequencies in different disease states.
Well-characterized samples from different vascular disease states (34 CREST, 41 MCTD, 24 scleroderma, and 35 Sjogren's syndrome) were scrutinized for the frequency of different ENAs when tested by AtheNA. All samples with negative results were retested by ELISA and were confirmed to be negative for all ENAs. The results of these determinations are detailed in Table 3.
Bead assay sensitivity of well-characterized SLE patients.
One hundred eighty-five clinically diagnosed and well-characterized
SLE patients were examined to determine the bead assay sensitivity.
The same samples were also tested by ELISA for ENAs and by Farr
for dsDNA. Table
4 details the results of this analysis. dsDNA
results by Farr were separated and analyzed with three different
cutoffs, 5 IU/ml, 10 IU/ml, and 15 IU/ml.
View this table:
[in this window]
[in a new window]
|
TABLE 4. Comparison of results for different autoantibodies detected by AtheNA Multi-Lyte and conventional EIA and Farr for 185 well-characterized lupus samples
|

DISCUSSION
Bead-based assays are fast becoming a mainstay in clinical laboratories.
The advantage of multiplexing combined with labor and reagent
cost savings makes these assays extremely attractive. The ability
to share a single platform to perform multiple assays is an
added bonus. Our goal for this study was to determine the clinical
value of these assays as it pertains to screening and detection
of autoantibodies and, furthermore, their ability to accurately
follow autoantibody changes during flares. Traditionally, screening
for autoantibodies has been performed by IFA on Hep-2 cells.
The sensitivity of this assay and its reliability have made
it the gold standard for screening purposes. On the other hand,
a low throughput, subjectivity, and the need for further testing
with single autoantibodies for identification of individual
autoantibodies have been the shortcomings for this assay. Screening
by ELISA has been suggested for overcoming throughput and technician
subjectivity, but lack of specificity has held it back from
replacing the IFA test. Bead-based assays now seem to fulfill
this shortcoming by their ability to multiplex. Concordance
values were in excess of 90% for four out of seven autoantibodies
between traditional methods and our bead-based assay (AtheNA
Multi-Lyte). The two autoantibodies with concordance levels
below 90% were RNP and SS/A. The low level of concordance with
RNP can be explained by the fact that the ELISA kit used for
detection of RNP detected the Sm/RNP combination, while the
AtheNA kit, due to its recombinant nature, detected only RNP.
This was more evident when we scrutinized 185 clinically diagnosed
SLE patients, where the difference in percentage of RNP detected
by the AtheNA and ELISA methods matched the percentage of Sm
detected by both. Similarly, the low concordance in SS/A can
be attributed to the differences in antigens used in the kits.
Further investigation from manufacturers allowed us to realize
that the AtheNA SS/A bead has a bias towards the 60-kDa fragment
of SS/A (Inova claims an equal concentration of both fragments),
although it includes both the 60- and the 52-kDa fragments.
This fact might also explain the discrepancy in the percentage
of positive SS/A samples in our characterized SLE patient group,
since lupus samples have been shown to have a higher bias towards
the 60-kDa fragment of SS/A (
13). This fact, though, did not
diminish the sensitivity of the AtheNA towards the Sjogren's
samples, which have been shown to have a higher bias towards
the 52-kDa fragment.
After establishing concordance with a proven methodology, we looked at false-positive and false-negative rates of the bead assay. For its use as a screening assay, it was important to determine if AtheNa produced a false, higher percentage of positive results (unspecific binding) than IFA when a normal population was scrutinized. Conversely, it was important to know the limits of the AtheNA assay when it came to incorrectly missing positive samples. For false positives, 7 out of 94 (7.4%) normal controls tested positive by AtheNA. Repeat testing with IFA showed that four out these seven samples had a titer of 1:40 and three had a titer of >1:160. These values are well within the established prevalence of positive ANA results in normal persons (20% to 30% with titers of >1:40 and 5% with titers of >1:160) (11). The other 87 samples were negative with both AtheNA and IFA. For false negatives, we specifically looked at well-characterized SLE samples, all of which were positive by IFA. Since SS/A and RNP are not specific to lupus and their incidence is well below the occurrence in normal persons, we only considered the false-positive results from the dsDNA in this group to be of significance. Considering the sensitivity of the Farr assay and its utility in monitoring flares, AtheNA's performance in missing only 6% of these samples was an adequate indicator of its sensitivity towards not missing positive samples. It is important to consider this result in the context of the samples tested. Almost all of the samples used for the study came from well-established SLE patients who have been through therapy for years. When we looked at our SLE group of patients positive by AtheNA, out of 185 samples we did not have a single sample that was positive for dsDNA alone. In fact, it is not very common to have new lupus patients with just elevated dsDNA antibodies, although this phenomenon is commonplace for patients who have been in therapy and are undergoing a flare. In this light and considering the high negative predictive value of the test, we believe that the test can be used as an effective screen for antinuclear antibodies.
To assess the utility of bead-based assays for reflex testing (conducting further tests to determine more detail) of positive ANA screens, we looked at previously diagnosed patients in order to have a better picture of the frequency of different autoantibodies detected by AtheNA. The results detailed in Table 3 show equal or better frequencies of autoantibodies associated with specific diseases. All of the negative samples by AtheNA were further confirmed by ELISA and were negative for all autoantibodies. In particular, we tested the positive samples from the SLE patients for a detailed comparison. The frequency percentages detailed in Table 4 show excellent similarity in all autoantibodies except SS/A, which we suspect is due to the differences in the percentages of antigens (52 kDa and 60 kDa) used by the different manufacturers. The other notable fact from these data was the difference in the percentages of dsDNA determined by Farr and AtheNA. It seems that the sensitivity of the AtheNA test is equivalent to 15 IU/ml by Farr.
It is important to note that the results from multiplex assays should always be interpreted within the clinical context, since the amount of information generated by such assays can act as a double-edged sword. As with any diagnostic test system, use of this information without proper clinical presentation, especially for screening tests, can lead to misdiagnosis and improper therapy, since positive ANA results are quite common in healthy individuals. It is also important to understand the fundamentals and limitations of these assays before using them. Patients with a positive ANA result by IFA will almost always have a positive result for ANA regardless of the fact that they are being treated or they are experiencing a flare (3, 5, 9). But, patients with a positive ANA result by AtheNA will not stay positive after they are treated. The difference, it is important to remember, comes from the fact that the AtheNA ANA test is composed of nine individual analytes towards specific autoantibodies which do subside while the patient is on therapy. On the other hand, the IFA test is comprised of all the components present in the HEp-2 cells. While specific autoantibodies subside due to therapy, there are still other components in the cell that do react with the patient's serum (16, 17). This new and innovative difference in assay design could lead to serious consequences in the real world if not understood precisely. For example, many patients do change physicians, and with an autoimmune history, the physician expects to have a positive ANA result from a patient's serum. But, if these patients were in treatment and the laboratory utilized a bead-based assay to determine the ANA status, there might be a negative ANA result, which could be very confusing to the physician. This phenomenon is very real and does cause problems and mistrust in a new test such as this one when misunderstood. The reality, on the other hand, is that there really is no need for an ANA screening test if a patient comes in with an extensive autoimmune history, since the physician already expects the test result for this patient to be positive. This is not to say that the bead-based assays are at a disadvantage for detecting ANA-positive patients who are being screened for the first time and that have not received any treatment. In such cases, the results of our study do show that the test is sensitive and specific enough to act as a screening test.
In conclusion, we found the multiplexed bead assay's performance to be satisfactory for screening, for reflex testing a positive result to determine the presence of specific autoantibodies, and for monitoring patients throughout the course of their disease by detecting flares.

ACKNOWLEDGMENTS
This work was partially supported by Inverness Medical Innovations
Inc.
We thank Robert I. Morris and Allan L. Metzger (RDL reference laboratory) and Mark J. Kopnitsky (Zeus Scientific Inc.) for their expert advice. We also thank John Sama for his continued help throughout the project. We are especially grateful to Armen Hartoonian for his help in data analysis and programming.

FOOTNOTES
* Corresponding author. Mailing address: Primex Clinical Laboratories, 16742 Stagg Street #120, Van Nuys, CA 91406. Phone: (310) 383-6059. Fax: (818) 779-1059. E-mail:
erik{at}primexlab.com 
Published ahead of print on 21 March 2007. 

REFERENCES
1 - Bernardini, S., M. Infantino, L. Bellincampi, M. Nuccetelli, A. Afeltra, R. Lori, A. Biroccio, A. Urbani, and G. Federici. 2004. Screening of antinuclear antibodies: comparison between enzyme immunoassay based on nuclear homogenates, purified or recombinant antigens and immunofluorescence assay. Clin. Chem. Lab. Med. 42:1155-1160.[CrossRef][Medline]
2 - Emlen, W., and L. O'Neil. 1997. Clinical significance of antinuclear antibodies: comparison of detection with immunofluorescence and enzyme-linked immunosorbent assays. Arthritis Rheum. 40:1612-1618.[Medline]
3 - Faria, A. C., K. S. Barcellos, and L. E. Andrade. 2005. Longitudinal fluctuation of antibodies to extractable nuclear antigens in systemic lupus erythematosus. J. Rheumatol. 32:1267-1272.[Abstract/Free Full Text]
4 - Farr, R. S. 1958. A quantitative immunochemical measure of the primary interaction between I BSA and antibody. J. Infect. Dis. 103:239-262.[Medline]
5 - Gill, J. M., A. M. Quisel, P. V. Rocca, and D. T. Walters. 2003. Diagnosis of systemic lupus erythematosus. Am. Fam. Physician 68:2179-2186.[Medline]
6 - Gniewek, R. A., C. Sandbulte, and P. C. Fox. 1997. Comparison of antinuclear antibody testing methods by ROC analysis with reference to disease diagnosis. Clin. Chem. 43:1987-1989.[Free Full Text]
7 - Gniewek, R. A., D. P. Stites, T. M. McHugh, J. F. Hilton, and M. Nakagawa. 1997. Comparison of antinuclear antibody testing methods: immunofluorescence assay versus enzyme immunoassay. Clin. Diagn. Lab. Immunol. 4:185-188.[Medline]
8 - Gonzalez, C., P. Guevara, I. Alarcon, M. Hernando, J. A. Navajo, and J. M. Gonzalez-Buitrago. 2002. Antinuclear antibodies (ANA) screening by enzyme immunoassay with nuclear HEp-2 cell extract and recombinant antigens: analytical and clinical evaluation. Clin. Biochem. 35:463-469.[CrossRef][Medline]
9 - Hahn, B. H. 2005. Systemic lupus erythematosus, p. 1960-1967. In D. Kasper (ed.), Harrison's principles of internal medicine, 16th ed. McGraw-Hill, New York, NY.
10 - Jaskowski, T. D., C. Schroder, T. B. Martins, C. L. Mouritsen, C. M. Litwin, and H. R. Hill. 1996. Screening for antinuclear antibodies by enzyme immunoassay. Am. J. Clin. Pathol. 105:468-473.[Medline]
11 - Kavanaugh, A., R. Tomar, J. Reveille, D. H. Solomon, and H. A. Homburger. 2000. Guidelines for clinical use of the antinuclear antibody test and tests for specific autoantibodies to nuclear antigens. Arch. Pathol. Lab. Med. 124:71-81.[Medline]
12 - Martins, T. B., R. Burlingame, C. A. von Muhlen, T. D. Jaskowski, C. M. Litwin, and H. R. Hill. 2004. Evaluation of multiplexed fluorescent microsphere immunoassay for detection of autoantibodies to nuclear antigens. Clin. Diagn. Lab. Immunol. 11:1054-1059.[CrossRef][Medline]
13 - Morozzi, G., F. Bellisai, A. Simpatico, G. Pucci, M. R. Bacarelli, V. Campanella, R. Marcolongo, and M. Galeazzi. 2000. Comparison of different methods for the detection of anti-Ro/SSA antibodies in connective tissue diseases. Clin. Exp. Rheumatol. 18:729-731.[Medline]
14 - Shovman, O., B. Gilburd, O. Barzilai, E. Shinar, B. Larida, G. Zandman-Goddard, S. R. Binder, and Y. Shoenfeld. 2005. Evaluation of the BioPlex(tm) 2200 ANA screen. Analysis of 510 healthy subjects: incidence of natural/predictive autoantibodies. Ann. N. Y. Acad. Sci. 1050:380-388.[CrossRef][Medline]
15 - Tan, E. M., T. E. Feltkamp, J. S. Smolen, B. Butcher, R. Dawkins, M. J. Fritzler, T. Gordon, J. A. Hardin, J. R. Kalden, R. G. Lahita, R. N. Maini, J. S. McDougal, N. F. Rothfield, R. J. Smeenk, Y. Takasaki, A. Wiik, M. R. Wilson, and J. A. Koziol. 1997. Range of antinuclear antibodies in "healthy" individuals. Arthritis Rheum. 40:1601-1611.[Medline]
16 - Wiik, A. S., T. P. Gordon, A. F. Kavanaugh, R. G. Lahita, W. Reeves, W. J. van Venrooij, M. R. Wilson, and M. Fritzler. 2004. Cutting edge diagnostics in rheumatology: the role of patients, clinicians, and laboratory scientists in optimizing the use of autoimmune serology. Arthritis Rheum. 51:291-298.[CrossRef][Medline]
17 - Yurasov, S., H. Wardemann, J. Hammersen, M. Tsuiji, E. Meffre, V. Pascual, and M. C. Nussenzweig. 2005. Defective B cell tolerance checkpoints in systemic lupus erythematosus. J. Exp. Med. 201:703-711.[Abstract/Free Full Text]
Clinical and Vaccine Immunology, May 2007, p. 505-509, Vol. 14, No. 5
1071-412X/07/$08.00+0 doi:10.1128/CVI.00034-07
Copyright © 2007, American Society for Microbiology. All Rights Reserved.