Clinical and Vaccine Immunology, May 2006, p. 598-601, Vol. 13, No. 5
1071-412X/06/$08.00+0 doi:10.1128/CVI.13.5.598-601.2006
Copyright © 2006, American Society for Microbiology. All Rights Reserved.
Division of Clinical Immunology, Department of Medical Technology, Faculty of Associated Medical Sciences, Chiang Mai University, Chiang Mai 50200, Thailand,1 Research Institute for Health Sciences, Chiang Mai University, Chiang Mai 50200, Thailand,2 Medical Biotechnology Unit, National Center for Genetic Engineering and Biotechnology, National Science and Technology Development Agency at the Faculty of Associated Medical Sciences, Chiang Mai University, Chiang Mai 50200, Thailand3
Received 25 December 2005/ Returned for modification 6 February 2006/ Accepted 6 March 2006
|
|
|---|
|
|
|---|
Recently a CD4 monoclonal antibody (MAb), named MT4, was generated in our laboratory (8). When white blood cells (WBC) (n = 4) were stained with MAb MT4, for all tested subjects, CD4+ lymphocytes showed strongly positive reactivity (Fig. 1B, donors 1 and 2). Monocytes from two out of four tested subjects were negative (Fig. 1B, donor 1), whereas the rest of the tested subjects showed very weak reactivity (Fig. 1B, donor 2). Granulocyte population in all tested subjects was negative (Fig. 1B, donors 1 and 2). In contrast, CD4 MAb clone Leu3a (Becton Dickinson) reacted to both CD4+ lymphocytes and monocytes but not to granulocytes in all tested subjects (Fig. 1C, donors 1 and 2). These results indicated that MT4 MAb is a unique CD4 MAb that recognizes CD4 protein on CD4+ lymphocytes but reacts weakly or not at all to CD4 on monocytes. We have no experimental results that bear on the question of why MAb MT4 fails to bind to monocytes. It is possible to speculate that MAb MT4 may be binding to a particular epitope of CD4 molecule which is exposed on lymphocytes but is sterically or conformationally obstructed on monocytes. It also can be that MT4 MAb has low affinity, so it fails to identify the low CD4 expression on monocytes.
![]() View larger version (32K): [in a new window] |
FIG. 1. Immunofluorescence analysis of the reactivities of CD4 MAbs MT4 and Leu3a with peripheral blood leukocytes. Whole-blood samples were stained with either irrelevant negative control MAb (A), CD4 MAb MT4 (B), or CD4 MAb Leu3a (C) purchased from Becton Dickinson and analyzed by lysed whole blood indirect immunofluorescence. Granularity (SSC) and PE fluorescence (FL2) were plotted to show the binding of the MAb to each leukocyte population. The fluorescence intensities of negative control MAb for all cell populations are marked by rectangles. Two subjects (donor #1 and donor #2) are shown as representative of four studied subjects.
|
Using the MT4-coated latex beads, a simple non-flow-cytometric method called the MT4 rosetting method was developed for enumeration of CD4+ lymphocytes in whole blood. The MT4 rosetting reagent consists of inert latex particles coated with MAb MT4. The reagent is used to manually enumerate the absolute CD4 count by visible light microscopy. To accomplish this, 15 µl of MT4 rosetting reagent was added to 50 µl of K3EDTA blood in a test tube and mixed gently for 5 min. Then, 15 µl of a blood-latex particle mixture was pipetted into another tube containing 50 µl of Turk's solution (glacial acetic acid, 3%; gentian violet, 0.012%) and mixed gently for 10 to 15 s. The sample was loaded into a hemacytometer chamber and then observed under a light microscope using a x40 objective. Cells having three or more latex particle attached to them (rosettes) were counted as CD4+ lymphocytes. A total of 300 WBC were counted, and the percentage of bead-rosetted CD4+ lymphocytes in WBC was calculated. The percentage of bead-rosetted CD4+ lymphocytes in WBC was used to calculate the absolute CD4+ cell count by multiplying the percentage of CD4+ lymphocytes by the total WBC count (cells/µl) and dividing by 100. To validate the rosetting method, numbers of CD4+ lymphocytes from 60 healthy and 140 HIV-infected individuals were determined by this method and compared with results obtained with standard flow cytometry. As shown in Fig. 2, a high correlation between the two methods was obtained (r = 0.941). The mean absolute CD4 counts ± standard deviations by the MT4 rosetting method and flow cytometry were 454 ± 336 and 492 ± 348 cells/µl, respectively. The accuracy of the MT4 rosetting method in identifying those individuals with fewer than 200 CD4+ lymphocytes/µl was evaluated, because this count has been incorporated as part of the Centers for Disease Control and Prevention case definition for AIDS to determine clinical levels of HIV disease. The MT4 rosetting method had a sensitivity of 90%, a specificity of 90%, and an accuracy of 90%. The positive and negative predictive values were 75% and 97%, respectively, using flow cytometry as the reference method. By Bland-Altman analysis (2) of all tested samples, the MT4 rosetting method yielded lower CD4+ lymphocyte counts than flow cytometry by a mean of 86 cells/µl, and the limits of agreement at 95% confidence interval were 323 to 151 cells/µl (Fig. 3A). Bland-Altman analysis of samples having CD4+ lymphocyte levels lower than 250 cells/µl indicated a mean difference of 45 cells/µl and limits of agreement at 95% confidence interval of 187 to 97 cells/µl (Fig. 3B).
![]() View larger version (14K): [in a new window] |
FIG. 2. Comparison of absolute numbers of CD4+ lymphocytes by flow cytometry and the MT4 rosetting method (n = 200).
|
![]() View larger version (16K): [in a new window] |
FIG. 3. Bland-Altman plot of the differences versus the means of CD4+ lymphocyte counts by flow cytometry and the MT4 rosetting method. Upper and lower horizontal lines mark the 95% confidence limits. (A) Bland-Altman analysis of all tested samples. (B) Bland-Altman analysis of samples having CD4+ lymphocytes at levels lower than 250 cells/µl.
|
The developed method is inexpensive, easy to perform, and reliable in identifying those individuals with CD4+ lymphocyte counts fewer than 200 cells/µl and has good correlation to flow cytometry. Therefore, the same CD4+ lymphocyte cutoff number suggested for the flow cytometric method by the Centers for Disease Control and Prevention can also be applied for the developed method. By this method, the only instruments needed to perform the assay are a hemacytometer and a light microscope, making this test especially useful in laboratories with minimal equipment and cost feasible in areas where budgets are very limited.
We are thankful to Sirinporn Intrawut and Umpa Yasamut for their technical assistance. We are also thankful to John McDermed, the Research Institute for Health Sciences, Chiang Mai University, for editorial assistance with the manuscript.
|
|
|---|