Clinical and Diagnostic Laboratory Immunology, April 2005, p. 513-519, Vol. 12, No. 4
1071-412X/05/$08.00+0 doi:10.1128/CDLI.12.4.513-519.2005
Copyright © 2005, American Society for Microbiology. All Rights Reserved.
Center for Comparative Medicine,1 Comparative Pathology Laboratory,2 Department of Pathology and Laboratory Medicine, University of California, Davis, CA 95616,4 The Jackson Laboratory, Bar Harbor, ME 046093
Received 18 October 2004/ Returned for modification 12 December 2004/ Accepted 11 February 2005
|
|
|---|
|
|
|---|
We have selected a relatively new technology, designated "multiple analyte profiling," from Luminex Corp. (Austin, TX), which allows simultaneous detection of multiple analytes in a small amount of sample (2). Up to 100 analytes can be measured in a single reaction. In addition, the multiple analyte profiling technology has been designed for high-throughput analysis. This method has been used to determine amounts of 15 different cytokines in small amounts of individual samples (2) as well as for detection of serum antibodies to multiple peptide epitopes (8), autoantigens (4), bacterial antigens (13, 14), and viral antigens (12).
The multiplex microbead assay is performed by the use of unique, fluorescently coded sets of polystyrene microbeads (5.6-µm diameter) (7, 15). A specific ratio of an orange and a red fluorophore is embedded within the matrix of a specific microbead set that enables identification of each bead set (7, 15). Microbead sets conjugated to known biomolecules are mixed and added to the test sample. Analytes in the sample react with biomolecules applied as a coating on the microbeads. Specific interactions are detected by a common reporter fluorochrome (e.g., phycoerythrin) conjugated to a secondary detection reagent. Thus, the multiplex microbead assay has a substantial advantage over the conventional immunoassay techniques with its ability to perform simultaneous detection of antibodies to several infectious agents in one reaction container.
This study focused on the development of a multiplex microbead assay for detection of antibodies to 10 infectious agents in mice. Two strains of mice (BALB/cj and C57BL/6j) were inoculated with several viruses and Mycoplasma pulmonis to develop a panel of single-positive test sera. Microbeads coated with antigens from these agents were highly specific and sensitive for multiplex serodetection. Comparisons with ELISA and IFA demonstrated that the multiplex microbead assay is as sensitive and specific as these conventional assays in mouse serodetection. Furthermore, a volume as small as 1 microliter of serum is sufficient to perform detection of antibodies to multiple infectious agents.
|
|
|---|
Cultures of viruses and M. pulmonis. GD7, MMV, and respiratory enteric orphan virus (Reo-3) were cultured in baby hamster kidney 21 (BHK-21) cells grown to confluency in Dulbecco's modified Eagle's medium containing 2 mM glutamine, penicillin/streptomycin, and 10% fetal bovine serum. Virus preparations were added to the confluent BHK-21 cells and allowed to adsorb to the cells at 37°C in the presence of 5% CO2, for 1 h on a rocking shaker. Cells were cultivated in Dulbecco's modified Eagle's medium, without antibiotics, and containing 2% heat-inactivated fetal bovine serum and 2 mM glutamine. Infected cell cultures were maintained until cytopathic effects were evident (48 to 120 h). Tissue culture medium containing virus was collected, clarified by centrifugation at 1,000 x g for 10 min, and passed through a 0.45-µm-pore-size filter. Virus was stored at 80°C. For isolation of MMV, infected cells were collected by centrifugation at 1,000 x g for 10 min. Cells were harvested by treatment with 0.1% trypsin in 2 mM EDTA and resuspended in 5 ml of medium per T75 flask. To release the virus, infected cells were subjected to three sequential freeze-and-thaw cycles. Cell debris was removed by centrifugation at 1,000 x g for 10 min. The supernatant was clarified and stored frozen at 80°C. Virus titers were determined by endpoint dilutions. Titers of GD7 and Reo-3 were measured on BHK-21 cells, and 324K cells were used to establish the titer of MMV.
M. pulmonis, obtained from the American Type Culture Collection (Manassas, VA), was cultured and isolated per supplier's instructions. Cells were collected by centrifugation of culture media at 3,700 x g and washed three times with PBS. Antigen was extracted by adding lysis buffer (1% Triton X-100 in PBS, containing protease inhibitor cocktail [Roche Diagnostics GmbH, Indianapolis, IN]) to the cell pellet. The contents were mixed by vortexing and incubated at 4°C for 20 min, and lysate was filtered through an 0.45-µm-pore-size filter. Total protein concentration was measured by the Bradford method (Bio-Rad, Hercules, CA). The lysate was aliquoted and stored at 80°C until used for coating microbeads.
Production of positive sera. Mice of two different strains (BALB/cj and C57BL/6j) were inoculated with different infectious agents (24 mice per agent, per mouse strain). All mice were SPF and were a gift from Jax West Division of the Jackson Laboratory (Bar Harbor, ME). Two sources of virus preparation were used for inoculations. One was a known amount (measured as total protein) of sucrose-gradient-purified virus from Advanced Biotechnologies Inc. These included Sendai virus, vaccinia virus, and NCDV. Alternatively, infectious agents were grown and titer was determined as described above. These included GD7, MMV, Reo-3, and M. pulmonis.
A few animals (4 to 12) from each mouse strain were initially test inoculated with various dilutions of different infectious agents prepared in Hanks minimal essential medium to estimate the minimal infectious dose. After the minimum dose was estimated, viruses were diluted to inoculate mice for the production of standard sera. The inoculation schedule is summarized in Table 1 for BALB/cj and in Table 2 for C57BL/6j mice. The natural route of infection was tested first. GD7, Reo-3, MMV, and Sendai virus were inoculated by intranasal route. For MHV, a cage of SPF mice was exposed to dirty bedding from MHV-infected mice. The MHV-infected mice that were used as a source of exposure were confirmed by ELISA to be seronegative for the other infectious agents including GD7, EDIM, ectromelia virus, MMV, mouse parvovirus (MPV), Sendai virus, M. pulmonis, and Reo-3. NCDV was inoculated by gastric lavage.
|
View this table: [in a new window] |
TABLE 1. Production of standard sera in BALB/cj mice
|
|
View this table: [in a new window] |
TABLE 2. Production of standard sera in C57BL/6j mice
|
After all mice seroconverted, as determined by commercial ELISA, they were euthanized with carbon dioxide and exsanguinated by cardiocentesis. Serum from all animals was stored at 80°C.
Sera from naturally infected mice positive for antibodies to multiple infectious agents. Mouse sera from various sources are routinely submitted to the Comparative Pathology Laboratory, University of California, Davis, for antibody analysis. As tested by commercial ELISA kits, serum samples from seven mice in a naturally infected mouse population were found to be positive for antibodies to 6 of the 10 prevalent infectious agents.
Antigen preparation for coating microbeads and ELISA plates. Viral and M. pulmonis antigens, for use in immunoassays, were prepared by 1:1 mixing of the infectious agent preparation with 0.5% Triton X-100 (Fisher Scientific, Fairlawn, NJ) and protease inhibitor cocktail (Roche) in PBS (pH 7.2). NCDV was further treated with 1 mM EDTA and sonication at low setting (Virtis Virsonic 60; Gardiner, NY). Vaccinia virus preparations were subjected to inactivation by psoralen treatment as previously described (18). Briefly, psoralen (Sigma Chemicals, St. Louis, MO) was added at a final concentration of 10 µg/ml to the vaccinia virus antigen prepared above (0.5 mg/ml) and subjected to UV irradiation for 10 min at 254 nm (Chromato Vue Transilluminator, model T5-15; UVP, Inc., San Gabriel, CA). All antigen preparations were stored frozen at 80°C. Prior to use, preparations were thawed and centrifuged in a microcentrifuge at 10,000 x g for 10 min to remove debris.
Coupling antigens to microbeads. Microbeads were purchased from Luminex Corp. (Austin, TX). Various antigen preparations were chemically cross-linked to the microbeads according to the manufacturer's instructions. Bead stock was resuspended by vortexing and treatment in a sonicator bath (15 to 30 s) (Branson 1510; Danbury, CT). An aliquot of 2.5 x 106 beads was removed and centrifuged at 21,000 x g for 2 min. Beads were resuspended in 80 µl of activation buffer (100 mM monobasic sodium phosphate; pH 6.3) by vortexing and sonication (15 to 30 s).
To activate the beads for cross-linking to proteins, 10 µl of 50-mg/ml sulfo-N-hydroxysulfosuccinamide (Pierce, Rockford, IL) was added, and beads were mixed by vortexing. Then 10 µl of 50-mg/ml 1-ethyl-3-[3-dimethylaminopropyl]carbodiimide (EDC; Pierce, Rockford, IL) was added, and beads were mixed again by vortexing. All incubations of beads were performed in the dark. The bead mixture was shaken on a rotary shaker at room temperature for 20 min and centrifuged at 21,000 x g for 2 min. Beads were washed twice with 250 µl of 50 mM morpholineethanesulfonic acid (MES; pH 6.0) buffer. To coat them with antigens, pelleted beads were resuspended in the relevant antigen preparation diluted in 50 mM MES (pH 6.0) buffer. Optimization of antigen concentration was performed by coating different microbead sets with a range of proteins between 10 and 200 µg/ml for each antigen. These microbead sets were then used to test against standard mouse sera, which are positive for antibodies to the relevant infectious agent. Microbead sets that provided the strongest specific signal for each antigen against the positive mouse sera were selected. The optimized protein concentration for each antigen was as follows: 25 µg/ml of vaccinia virus, 100 µg/ml of NCDV, 25 µg/ml of M. pulmonis, 100 µg/ml of MHV, 25 µg/ml of Sendai virus, 100 µg/ml of GD7, 50 µg/ml of CMV, and 100 µg/ml of MMV.
Recombinant protein rVP2, a specific antigen for MPV serodetection (9), was obtained from the Research Animal Diagnostic Laboratory (University of Missouri, Columbia, MO). Optimal concentration of rVP2 for coating on the beads was determined to be 3.3 µg/ml. For Reo-3, infected cell lysate was used at 330 µg/ml, and a separate bead set was coated with uninfected cell lysate (Reo-3 sham) at the same concentration.
Bead sets were also coated with biotin-conjugated goat immunoglobulin G (IgG; 100 µg/ml), a positive control protein for reaction with streptavidin conjugated to R-phycoerythrin (streptavidin-phycoerythrin), and bovine serum albumin (BSA;100 µg/ml) as a negative control protein (Pierce, Rockford, IL). For coupling, mixtures of activated beads and proteins were incubated by shaking on a rocker for 2 h at room temperature. After coating with proteins, beads were washed twice with 250 µl of wash buffer (0.1% Tween 20 in PBS, pH 7.4), resuspended in 250 µl of blocking buffer (1% BSA; 0.1% Tween 20 in PBS, pH 7.4; 0.05% sodium azide), and shaken on a rocker at room temperature for 30 min. After blocking, beads were resuspended in 1 ml of blocking buffer and stored at 4°C.
Multiplex detection of antibodies in mouse serum. Immunoreactions were done in 96-well, filter-bottomed plates designed for high-throughput separations (1.2-µm-pore-size MultiScreen; Millipore Corporation, Bedford, MA). Typically, 2,000 beads for each individual bead set, coated with a specific antigen, were added per well. For example, for a 13-plex assay, 2,000 beads from each set coated with an antigen were mixed to provide a total of 26,000 beads per well. Mouse serum was diluted 1:250 in 5% BLOTTO (Pierce, Rockford, IL), and 50 µl of this diluted serum was mixed with the 13-plex bead mixture per well. The mixture of beads and serum was then incubated on a shaker for 1 h at room temperature. After incubation, liquid was drained from the bottom of the plate in a vacuum manifold designed to hold 96-well plates (Millipore Corporation, Bedford, MA). The beads were washed two times by adding 100 µl of wash buffer per well and draining under vacuum. For detection of mouse IgG, biotinylated horse anti-mouse IgG was used (Vector Laboratories, Burlingame, CA) at a 1:1,000 dilution in BLOTTO, and 100 µl was added per well. Beads were mixed as before and incubated at room temperature for 30 min. Following incubation with the secondary antibody, beads were washed three times. To detect biotinylated IgG, 100 µl of streptavidin-phycoerythrin (CalTag, Burlingame, CA) was added at a dilution of 1:500 in wash buffer. The content of each well was mixed and incubated at room temperature for 15 min. Beads were washed once with wash buffer, resuspended in 100 µl of wash buffer per well, and analyzed in the Luminex-100 instrument equipped with XY-Platform for automatic reading of a 96-well plate.
Luminex-100 instrument operation and data analysis. The Luminex-100 instrument was manufactured by Luminex Corp. and purchased from Upstate USA Inc. (Lake Placid, NY). The instrument was used at default settings, set by the manufacturer for routine applications, as directed by the user's manual. Data were acquired by Luminex Data Collection software (version 1.7). This software package allowed routine operation of the instrument and data acquisition. Calibration beads, supplied by the manufacturer, were used to adjust instrument settings for bead set identification and for the detection of reporter (phycoerythrin). Events were gated to exclude doublets and other aggregates. A minimum of 100 independent, gated events were acquired for each bead set. A ratio of median fluorescence intensity (MFI) of antigen-coated bead sets to MFI of BSA-coated beads in each reaction was calculated. This ratio, designated as "signal," was used as a measure of antibody detection. After acquisition by Luminex software, the data were further processed by Microsoft Excel software. Background or normal serum reactivity for the antigen-coated bead set was determined. For each bead set, an average of signals from quadruplicate wells of pooled normal mouse sera was calculated, and three times the standard deviation value was added to the average. For routine analysis, each experiment also contained a positive-control serum for all the infectious agents. Samples were analyzed in duplicate, and a minimum of two separate multiplex immunoassay experiments was performed. Sera were designated positive for antibodies to an infectious agent if the signal from the relevant antigen-coated bead set was greater than the background.
IFA. IFA analysis of serum samples was performed by Charles River Laboratories (Wilmington, MA).
ELISA. Two types of ELISA systems were used. One system consisted of preconfigured commercial kits for the serodetection of different mouse infectious agents. These ELISA kits, purchased from Charles River Laboratories (Wilmington, MA), were used to test mouse sera according to the manufacturer's instructions. The other ELISA system was developed with the purified antigens that were also used in the development of the multiplex microbead assay. This ELISA was used for routine testing of the antigens prior to coating on the microbeads. Immulon 4 HBX, flat-bottomed microtiter plates (ThermoLabsystems, Franklin, MA) were coated with viral preparations in the following optimized protein concentrations: Sendai virus was used at 0.1 µg/ml, MHV at 0.01 µg/ml, GD7 at 1.0 µg/ml, MMV at 3.0 µg/ml, MCMV at 1.0 µg/ml, NCDV at 0.01 µg/ml, and vaccinia virus at 0.3 µg/ml. MPV recombinant antigen was used at 0.1 µg/ml. Reo-3-infected and uninfected cell lysates were used at 50.0 µg/ml each, and M. pulmonis lysate was used at a 10.0-µg/ml total protein concentration. Antigens were diluted to a final optimized concentration in coating buffer: Hanks balanced salt solution containing 0.375% sodium bicarbonate. Antigens were added to plates (100 µl/well) and incubated overnight at 4°C. Plates were washed with wash buffer (0.1% Tween 20 in PBS, pH 7.4) and blocked with BLOTTO (5% nonfat powdered milk in wash buffer) for 2 h at room temperature. To each well, 100 µl of mouse serum, diluted 1:250 in BLOTTO, was added. Plates were allowed to incubate at room temperature for 1 h and washed again with wash buffer. Detection antibody, biotinylated horse anti-mouse IgG (Vector Laboratories, Burlingame, CA), was diluted 1:10,000 in wash buffer and added to the ELISA plate at 100 µl/well. Plates were incubated with the detection antibody for 30 min at room temperature. After washing, 100 µl of Vectastain ABC detection reagent, containing horseradish peroxidase H (Vector Laboratories), was added to each well. Plates were washed and developed by adding 100 µl/well of TMB color development reagent (Sigma Chemicals, St. Louis, MO). Plates were developed at room temperature for 10 to 12 min, and the reaction was stopped by adding 50 µl/well of 1 M sulfuric acid. Color development was measured at 450 nm in the ELISA plate reader (Bio-Rad, Hercules, CA).
|
|
|---|
Standard serum production in BALB/cj mice (Table 1). Five viruses, Sendai virus, GD7, Reo-3, MHV and NCDV, produced infections in BALB/cj mice inoculated by the natural route. However, more than one dose was usually required to induce seroconversion in most of the animals. For NCDV, a majority of the mice were inoculated by the i.p. route, which resulted in 100% seroconversion (Table 1). For MVM, M. pulmonis, and vaccinia virus, mice were inoculated by the i.p. route alone. These three agents failed to generate antibodies by other routes of exposure (Table 1).
Standard serum production in C57BL/6j mice (Table 2). The course of infection and antibody generation in the C57BL/6j mice was similar to that described above for BALB/cj mice. However, it appears that inoculations in C57BL/6j mice by Sendai virus or GD7 virus were more successful in generating antibody responses than in BALB/cj mice. For example, 23 mice inoculated with Sendai virus and 22 mice inoculated with GD7 by the intranasal route became seropositive after only one inoculation. In contrast, most BALB/cj exposed to Sendai virus or GD7 virus by routes mimicking transmission under colony conditions required reinoculation for the induction of antibody responses.
Analysis of single-positive mouse sera by multiplex microbead assay. Serum samples, representative of individual groups of mice inoculated with specific infectious agents, were analyzed by multiplex microbead assay. Individual, antigen-coated and control microbeads were mixed to produce the 13-plex bead mixture (see Materials and Methods). Sera were diluted 1:250, unless stated otherwise, and incubated with the 13-plex microbead mixture. The results of the multiplex assay, presented in Table 3, show that specific antibodies to different infectious agents were detected with high sensitivity by the antigen-coated bead sets. Importantly, these beads did not display a significant nonspecific reaction to mouse sera, compared to background signal from normal mouse sera. Thus, these findings demonstrate that the multiplex microbead assay is highly specific and sensitive for the detection of antibodies to the panel of 10 mouse infectious agents.
|
View this table: [in a new window] |
TABLE 3. Antibody detection in known singly positive sera by multiplex microbead assaya
|
|
View this table: [in a new window] |
TABLE 4. Multiplex microbead assay on mixed, singly positive mouse sera from a
|
|
View this table: [in a new window] |
TABLE 5. Multiplex assay for serodetection in field samples of micea
|
|
|
|---|
To develop the multiplex microbead assay system, we produced mouse sera positive for antibodies to individual infectious agents (Tables 1 and 2). Attempts were made to inoculate these mice by routes mimicking natural transmission. Because our goal was to develop multiplex serodetection with wide applicability to mouse strains that range in magnitude of antibody responses, the BALB/cj and C57BL/6j strains were used to produce monospecific sera. Sera from these experimentally inoculated mice were critical for unambiguous evaluation of sensitivity and specificity of the multiplex microbead assay. Mouse sera obtained from commercial sources are generated by inoculating mice with a mixture of several antigens. These sera may be positive for antibodies to multiple agents but are provided as positive for only one agent; antibodies to any other agents, if present, are not specified. Such sera may be useful to test the performance of the conventional immunoassays such as ELISA, IFA, and Western blotting. However, these commercial sera are unsuitable for use in the multiplex microbead assay. This complication was further exacerbated by lot-to-lot variation of the sera, resulting in high titers against a different set of antigens in each lot. Accordingly, the monospecific sera produced in this project will be valuable reference material for future assay development and validation, particularly as the multiplex microbead assay is expanded for serodiagnosis of additional infectious agents.
The multiplex serodiagnostic assay was tested by using sera from animals containing antibodies to single infectious agents. As shown in Table 3, microbeads coated with individual antigens reacted to mouse sera in a specific manner. The data also demonstrated that the microbeads did not display nonspecific reactivity to sera containing irrelevant antibodies. The specificity of the multiplex microbead assay was further tested in simultaneous serodetection of multiple agents in a single reaction. Two experiments were performed: (i) standard positive sera were mixed and reacted against the antigen-coated bead sets (Table 4) and (ii) field sera from mice infected with several agents were obtained from a mouse facility and tested (Table 5). Both experiments clearly demonstrated a high degree of sensitivity and specificity for the multiplex microbead assay. In Table 4, where positive sera from several mice were mixed, minor nonspecific reactivity of the antigen-coated beads was observed, e.g., in the serum mixture lacking MMV-positive serum, resulting in a decrease in the signal-to-noise ratio. These results may reflect potential interactions and interference among sera from different animals. However, any such inconsistencies were resolved in sera from animals infected with multiple agents, yielding results that were clearly interpretable (Table 5). Similar types of minor interferences have also been observed in multiplex immunoassay in a mixture of individual sera from rhesus macaques positive for antibodies to a single infectious agent. Again, such interferences were not observed in individual rhesus macaque sera positive for antibodies to multiple infectious agents (I. Khan and P. Luciw, unpublished data).
Clinical validation of the multiplex microbead assay will require analysis of a large number of field serum samples that are also tested by conventional immunoassays, which detect antibodies to one infectious agent at one time. Accordingly, several thousand samples will be analyzed to establish statistically significant levels of sensitivity and specificity. In addition, a high-throughput format of the multiplex microbead assay is currently under evaluation to facilitate validation.
The increased efficiency of the multiplex microbead immunoassay and the requirement for only one microliter of serum will promote direct testing of individual mice. In current practice for health maintenance of SPF colonies, indirect testing by the use of sentinel mice may have certain disadvantages. For example, a recent report showed that some sentinel mice used in serosurveillance were infected with MPV at the time of arrival from the rodent vendor and were the source of MPV infection in the colony animals (16). Thus, direct testing of mice may improve the accuracy of health monitoring in mouse colonies. Because the multiplex microbead assay requires only 1 microliter of sample, this method is ideally suited for serosurveillance of mouse colonies in general, and in particular of more valuable mice, such as genetically engineered lines. In addition, the ability to detect and measure multiple analytes in a small volume of serum (or plasma) allows for analysis of samples collected in the live phase of a research protocol.
This work was supported by PHS grant RR14034 to S. W. Barthold from the National Center for Research Resources, NIH.
|
|
|---|