Previous Article | Next Article ![]()
Clinical and Diagnostic Laboratory Immunology, March 2004, p. 399-405, Vol. 11, No. 2
1071-412X/04/$08.00+0 DOI: 10.1128/CDLI.11.2.399-405.2004
Copyright © 2004, American Society for Microbiology. All Rights Reserved.
and Allan Hildesheim1
Division of Cancer Epidemiology and Genetics, National Cancer Institute, NIH, DHHS, Bethesda, Maryland,1 Proyecto Epidemiológico Guanacaste, San José, Costa Rica,2 Departments of Obstetrics, Gynecology, and Reproductive Biology, Fearing Research Laboratory, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts3
Received 24 September 2003/ Returned for modification 3 December 2003/ Accepted 8 December 2003
|
|
|---|
), granulocyte-macrophage colony-stimulating factor (GM-CSF), immunoglobulin A (IgA), or IgG, and sponges were extracted and evaluated for total recovery by enzyme-linked immunosorbent assay (ELISA). There was excellent (>75%) recovery for all immune markers from all three devices except for IL-6, which was poorly recovered (<60%) for all sponge types, IFN-
, which was poorly recovered from both Weck-Cel and Ultracell sponges but was completely recovered from Merocel sponges, and IL-4, which was poorly recovered from Weck-Cel sponges but was completely recovered from Ultracell or Merocel sponges. We then compared the absolute recovery of selected markers (IL-10, IL-12, IgG, and IgA) from cervical secretion specimens collected from women using each type of sponge. There were no significant differences in the recoveries of IL-10, IL-12, and IgG from cervical specimens collected by any type of ophthalmic sponge, but there was reduced IgA recovery from Merocel sponges. However, the variability in these measurements attributable to sponge types (1 to 3%) was much less than was attributable to individuals (45 to 72%), suggesting that differences in sponge type contribute only in a minor way to these measurements. We infer from our data that the three collection devices are adequate for the measurements of IL-1ß, IL-2, IL-5, IL-12, IL-15, IL-18, and IgG. Merocel may be a better ophthalmic sponge for the collection of cervical secretions and measurements of IL-4, IL-8, IL-10, GM-CSF, and IFN-
, but our data from clinical specimens, not in vitro-loaded sponges, suggested the possibility of reduced recovery of IgA. These findings require confirmation. |
|
|---|
), were poorly recovered from in vitro loading of these sponges (16). The reason for poor recovery of these cytokines is uncertain, but one possible explanation is that specific immune markers irreversibly bind to the absorbent material of this sponge.
To address whether the composition of the sponge might improve the accuracy of measuring immune markers in cervical secretions, we compared the recoveries of immune markers from Weck-Cel ophthalmic sponges (composed of cellulose), Ultracell (Ultracell Medical Technologies, North Stonington, Conn.) ophthalmic sponges (composed of polyvinyl alcohol), and Merocel (XOMED Surgical Products, Inc.) ophthalmic sponges (composed of hydroxylated polyvinyl acetate) loaded in vitro and in vivo. For our in vitro investigations, we selected a broad spectrum of immune biomarkers to load onto the sponges that represent T-cell helper 1 (Th1) and 2 (Th2) responses and included the aforementioned problematic cytokines, IL-4 and IFN-
. For our in vivo investigations, the number of markers we can explore from a sponge is limited; we thus selected IgA and IgG recovery from sponges loaded with cervical secretions to identify the optimal sponge for our vaccine work as well as two representative cytokines, IL-10 and IL-12, that we typically measure in cervical secretions from our natural history studies of HPV (7, 12).
|
|
|---|
, or granulocyte-macrophage colony-stimulating factor, from commercial enzyme-linked immunosorbent assay (ELISA) kits (BioSource International, Camarillo, Calif.) or with commercially available IgG and IgA (The Binding Site, Birmingham, United Kingdom). The lyophilized standards were reconstituted with the diluent provided with the kit. The standard was then further diluted so that 100 µl of the diluted immune marker absorbed by sponges would have a final concentration in the extraction volume equal to the median of the standard curve if 100% of the material was recovered. For each immune marker, five spears of each type were spiked with diluted immune marker, allowed to absorb for 10 min, and then extracted. As a background control, an additional spear was loaded with extraction buffer alone (no immune marker) and tested, and the resulting value was subtracted from test values. As a positive control, a volume of the diluted immune marker was added directly to the extraction buffer and tested, and the result was compared to the expected 100% value. All ELISAs were run on the same day the spears were extracted to minimize degradation of the immune marker. All samples were run in duplicate in the ELISAs, and the average of these two values was used. Extraction of ophthalmic sponges. Sponges were extracted according to a previously described protocol (16). Briefly, each sponge was weighed to estimate the volume of secretions absorbed onto the sponge. The sponge was then equilibrated in 300 µl of extraction buffer (phosphate-buffered saline, 0.25 M NaCl, 0.1 mg of aprotinin per ml, and 0.001% sodium azide) for 30 min at 4°C. Sponges were then centrifuged at 16,000 x g in a Spin-x centrifuge filter unit (Costar, Cambridge, Mass.) to separate extracted samples from the sponge matrix. A second wash of the sponge with 300 µl of extraction buffer was followed by immediate centrifugation in the Spin-X tube. The two extraction volumes were combined, 5 µl was removed for hemoglobin testing by Hemastix (VWR International, West Chester, Pa.), and then 4 µl of fetal calf serum was added to the remaining specimen. A dilution factor was calculated based on the estimated volume and dry weight of the spear as previously described (8): [(x - y) + 0.6 g of buffer]/(x - y), where x equals the weight of the sponge after collection and y is the mean weight of the dry spear, based on independently weighing six spears.
Collection and testing of clinical specimens. Cervical secretions were collected from 90 consenting women scheduled for a colposcopy in the context of an National Cancer Institute (NCI)-sponsored, NCI- and local institutional review board-approved population-based cohort study of HPV and cervical neoplasia established in Guanacaste, Costa Rica, from 1993 to 1994 (9, 10). Women were scheduled for colposcopy examinations either for (i) an exiting colposcopic inspection as a participant in the natural history study; (ii) a follow-up visit after the exit colposcopy; (iii) an evaluation and, if necessary, treatment of high-grade cervical neoplasia; or (iv) a follow-up visit after a loop electrosurgical excision procedure of the transformation zone to treat high-grade cervical neoplastic lesions. Specimens were collected every day from the first 10 consenting women attending their colposcopic appointment plus all women with indications of high-grade cervical neoplasia or from women from whom we had originally collected cervical secretions until 90 participants were accrued. Participants were randomly assigned one of six combinations of paired sponges (Merocel-Weck-Cel, Merocel-Ultracell, Ultracell-Merocel, Ultracell-Weck-Cel, Weck-Cel-Merocel, and Weck-Cel-Ultracell). Cervical secretions were collected as paired specimens as previously described (12). Specimens were collected by a single practitioner (L. A. Morera) by placing the sponge portion of the ophthalmic device within the os of the cervix for 30 seconds, removing the device from the patient, and immediately placing it into a cryovial and into liquid nitrogen. The second specimen was collected in a similar manner. Since we began collecting cervical secretions in 1996, these collections have been well tolerated, with no reports of adverse events or noncompliance, and the same was true for the three devices evaluated in this study.
The same commercial ELISA kits used to assay the in vitro-loaded sponges were used to assay extracted clinical specimens for IL-2, IL-10, IL-12, IFN-
, IgG, and IgA. Specimens, extracted in the same fashion as the in vitro-loaded sponges, were tested for IL-2, IFN-
, IgG, and IgA by using undiluted volumes of 200 µl, 200 µl, 10 µl (diluted 1:1,000), and 10 µl (diluted 1:1,000), respectively. Specimens were tested for IL-10 and IL-12 by using 80-µl volumes diluted 1:1.5 with the assay diluent. We did not evaluate IL-4 or the other immune markers primarily because of the limited number of markers that could be measured from a single specimen and the need to test whether IL-10, IL-12, IgG, and IgA, our standard markers of Th1 and Th2 responses, were recovered equally well from the three sponge types.
Analysis. Total amounts of immune markers recovered from in vitro-loaded sponges were calculated by multiplying the final extracted volumes and the concentrations from the ELISA. The total amount recovered was divided by the total amount loaded onto the sponge to calculate the mean percent recovery and 95% confidence interval. Differences in percent recovery between sponges were tested for statistical significance by a Z test.
To assess the randomization of women into the six study groups, we evaluated the differences between groups for a small number of sociodemographic, reproductive, and sexual variables (age at visit, use of oral contraceptives, lifetime number of pregnancies [<4 or
4], years of education, age at first pregnancy, and age at first sexual intercourse) collected through an interviewer-administered questionnaire. Statistical differences in median values for continuous variables and in categorical variables were assessed by a Kruskal-Wallis test, a nonparametric analysis of variance statistic, and a Pearson
2 test, respectively.
The measurements of the concentrations of immune markers extracted from each type of ophthalmic sponge were corrected for dilution as the result of extraction and compared by the Kruskal-Wallis test. First and second sponges were evaluated separately and then combined in a separate analysis. For the analysis of IL-10 and IL-12 concentrations, specimens with values below the limits of detection (0.2 pg/ml for IL-10 and 0.8 pg/ml for IL-12) were assigned a value of one-half the limit of detection (14 first sponges and 16 second sponges for IL-10; 2 first sponges and 1 second sponge for IL-12). Only three first spears and seven second spears had detectable levels of IFN-
(limit of detection, 4 pg/ml), and all IL-2 measures were below the limits of detection (5 pg/ml). Therefore, neither immune marker could be evaluated from cervical secretions, and only IgA, IgG, IL-10, and IL-12 levels in cervical secretions were evaluated.
Variance component analyses using mixed models (random and fixed effects) were performed (12, 17, 18) to explore the relationships of covariates with the levels of the IgA, IgG, IL-10, and IL-12 in cervical secretions. The type of sponge (Merocel, Ultracell, or Weck-Cel), the order of collection (first or second sponge), the volume of the secretions (quartile categories of <14.5, 14.5 to 22.8, 22.9 to 43.4, and
43.5 µl), and the level of hemoglobin in the secretions (categories of 1 to 4, 5 to 6, and 7, as measured by the Hemastix) were treated as fixed effects. The individuals in the study were treated as random effects. Boxcox transformations [Y = (y
- 1)/
, in which y is the measured value and
is the Boxcox coefficient] (1) of IgA, IgG, IL-10, and IL-12 concentrations were used to normalize the data (NB, logarithmic transformation was not sufficient to create a normal distribution of values); values of
for IgA, IgG, IL-10, and IL-12 were 0.0606572, 0.0921632, 0.1238986, and 0.1822423, respectively. In univariate analyses, hemoglobin levels were not associated with any of these immune markers and, thus, were excluded from additional analyses. To estimate the relative contribution of variance from individuals versus the type of the sponge, a model was run in which sponge type was also treated as a random effect. Adjusted means were calculated on transformed values by using the least squared means command of the SAS/STAT software (version 8; SAS Institute, Cary, N.C.), and the values were untransformed. In multivariate models, education and age did not appreciably alter the estimates of adjusted mean concentrations of the immune markers and thus were excluded from the final models.
Finally, to examine the relative performance and determine if there were systematic errors associated with any sponge type, we examined the correlation of measurements from the first sponge and the combination of both second sponges (e.g., a Merocel first sponge versus an Ultracell or Weck-Cel second sponge) by using scatter plots and calculating Spearman correlations (
).
|
|
|---|
, almost no IL-4 (5%), and less than 75% of granulocyte-macrophage colony-stimulating factor, IL-6, IL-8, and IL-10 were recovered from in vitro-loaded Weck-Cel sponges (Fig. 1); all other analytes were quantitatively (>75%) recovered. Only 15% of IFN-
and less than 75% of IL-6, IL-8, and IL-10 were recovered from in vitro-loaded Ultracell sponges. Only 15% of IFN-
and less than 75% of IL-6, IL-8, and IL-10 were recovered from Ultracell sponges; all other analytes were quantitatively (>75%) recovered. All analytes were quantitatively (>75%) recovered from in vitro-loaded Merocel sponges except for IL-6 (58%).
![]() View larger version (66K): [in a new window] |
FIG. 1. Mean percentages of recovery from different ophthalmic sponge types (black columns, Weck-Cel; gray columns, Ultracell; white columns, Merocell) loaded in vitro with known amounts of immune markers. Bars indicate 95% confidence intervals. Statistically significant differences (P < 0.05) are indicated by the following symbols: *, Weck-Cel < Ultracell; , Weck-Cel < Ultracell and Merocell; , Weck-Cel and Ultracell < Merocel; #, Weck-Cel < Ultracell < Merocel.
|
levels below the limits of detection; thus, these cytokines could not be evaluated for recovery from clinical specimens. The mean, median, and range of the estimated volumes of secretions for the first sponge were 46, 32, and 1 to 240 µl, respectively, and for the second sponge were 29, 18, and 1 to 240 µl, respectively. The estimated median volumes collected by Weck-Cel sponges (first sponge, 42 µl; second sponge, 25 µl) were not significantly greater than for Ultracell sponges (first sponge, 29 µl; second sponge, 17 µl) or for Merocel sponges (first sponge, 32 µl; second sponge, 16 µl) (P = 0.09, first sponge; P = 0.06, second sponge). |
View this table: [in a new window] |
TABLE 1. Characteristics of women randomized to the six study groups
|
|
View this table: [in a new window] |
TABLE 2. Estimated median values of IL-10, IL-12, IgG, and IgA for cervical secretions extracted from different ophthalmic spongesa
|
|
View this table: [in a new window] |
TABLE 3. Adjusted means of IgA, IgG, IL-10, and IL-12 levels from cervical secretions collected with ophthalmic spongesa
|
![]() ![]() View larger version (37K): [in a new window] |
FIG. 2. Correlations between the first sponge and second sponge (Merocel versus Ultracell or Weck-Cel, Weck-Cel versus Merocel or Ultracell, and Ultracell versus Merocel or Weck-Cel, respectively) for IgA, IgG, IL-10, and IL-10. Spearman correlations are shown for the first sponge versus both types of second sponges.
|
|
|
|---|
may be poorly recovered from Weck-Cel ophthalmic sponges (16) and demonstrated that immune markers may be more reliably recovered from in vitro-loaded Ultracell and Merocel ophthalmic sponges. In particular, recovery from Merocel sponges, which are composed of hydroxylated polyvinyl acetate, was nearly 100% for all tested cytokines in the in vitro-loading experiment, except for IL-6, whose levels of recovery were equally poor from all sponges. Because of the complex nature of cervical secretions compared to in vitro spiking of the collection devices, we sought to verify whether there were appreciable differences between the levels of detection of immune markers from clinical specimens. Despite the equally high recovery of IgA from each sponge type in our in vitro experiments, there was some evidence of reduced IgA recovery from clinical specimens when Merocel sponges (and Weck-Cel sponges) were used for collection. Although we randomized the women to different groups and attempted to adjust for any differences between groups (using both order and type), the groups were not perfectly balanced (e.g., as evident by the differences in age), and reduced IgA recovered from Merocel sponges could also be due to chance differences between groups. This observation is preliminary, however, and warrants confirmation. Moreover, differences in these measurements due to sponge types are relatively minor compared to the large differences between individuals, the latter being consistent with previous observations (7, 12). The large variability in genital tract immune biomarker concentrations between women is due in part to many possible determinants; for IL-10 and IL-12, they may include time of the menstrual cycle, low secretion volume, macro levels of heme contamination, high vaginal pH, numbers of lifetime sex partners, frequency of sexual intercourse, present oral contraceptive use, and smoking status (7).
In this study, there was also a nonsignificant tendency for higher IL-10 and IL-12 concentrations from Merocel sponges than from the other types. However, we note that there were no apparent systematic errors in the use of any of these sponges (Fig. 2), although there was the expected tendency of poorer correlation (greater scatter) as the measurements approached limits of detection, as best illustrated by IL-10 scatter plots. Thus, we anticipate that any of the sponges will measure relative differences in any immune marker equally well, provided that losses in immune marker recovery from the sponge do not reduce concentrations near or below the limits of detection.
One limitation of our study was that we were not able to evaluate the recovery of IFN-
from any of the ophthalmic sponges used to collect cervical secretions due to mostly undetectable levels, and thus we were unable to confirm these differences in vivo. IL-2 was also not detectable from clinically collected specimens. IL-2 is detectable in adolescent women (5) but levels are substantially lower in older women (P. Crowley-Nowick, unpublished observations); the median age in this study was 43 years. Our study was designed primarily to detect the large differences observed for IFN-
in vitro (post hoc analysis indicates a
80% power to detect a 90% reduced recovery), and thus subtle differences in recovery would not be detectable.
A second limitation is that the small volumes of cervical secretions recovered (
50 µl) and the low concentrations of some immune markers after extraction restricted the number of immune markers that could be measured by ELISA in this validation study. Specifically, we did not examine IL-4, which appeared to be problematic for Weck-Cel sponges as determined from our in vitro measurements. We are now exploring other assay systems that permit multiple measurements from the same aliquot of cervical secretions and thereby circumvent this limitation (3).
We conclude that there may be some advantages to using Merocel ophthalmic sponges instead of Weck-Cel or Ultracell sponges. However, further studies are needed. The unexplained tendency of poorer IgA recovery from clinical specimens collected with Merocel sponges warrants caution, and it may be necessary to select sponges based on specific study objectives. For example, if the lower IgA recovery from Merocel sponges is confirmed, it may be less appropriate to use this sponge to monitor anti-HPV IgA responses to vaccination. We again note that potential differences between sponges are small compared to the differences between women, which are likely due to a number of factors (e.g., time of the menstrual cycle) that contribute to differences in genital tract physiology. Finally, optimizing specimen collection for measurements of genital tract immune responses may be important to human studies such as vaccine trials for protection against sexually transmitted infections.
We gratefully acknowledge the IMS group (Rockville, Md.) of Julie Buckland and John Schussler. We offer special recognition for the excellent work of the study staff in Costa Rica. We also acknowledge the collaboration of health authorities in Costa Rica for their enthusiastic support of this project and of the outreach workers of the Ministry of Health of Costa Rica, who carried out the population census, for their dedication to the health of the people of Guanacaste. Finally, we thank Lisa M. McShane (Biometric Research Branch, NCI, NIH, DHHS) for her statistical expertise.