Cytokines are involved in numerous immunological functions, having both antagonistic and synergetic effects on many different cell types as well as enhancing the production of other cytokines. The optimal manner in which to correlate a specific disease process with changes in cytokine concentrations requires analyzing individual samples for multiple cytokines. The enzyme-linked immunosorbent assay (ELISA) is the most commonly reported method for the quantitation of secreted cytokines. These assays, however, must all be run individually, which in turn requires substantially more reagents, more technician time, and a larger sample volume. Employing a multiplexed fluorescent microsphere immunoassay system (Luminex 100; Luminex Corp., Austin, Tex.), we have developed a sandwich capture assay to simultaneously assess the production of nine different cytokines (interleukin-1ß [IL-1ß], IL-2, IL-4, IL-6, IL-8, IL-10, IL-12 p70, gamma interferon [IFN-
], and tumor necrosis factor alpha [TNF-
]) and one cytokine receptor (IL-2 soluble receptor alpha [IL-2r]) in human serum. Also included in the multiplexed assay were internal controls for detecting interfering levels of heterophile antibodies.
Heterophile antibodies are a well-recognized cause of interference in immunoassays (4, 12, 14, 19) and are present in 5 to 40% of normal blood donors (7, 8, 11). Heterophile antibodies (antibodies which cross phyla in their reactivity) are produced against poorly defined antigens and generally show weak avidity and are multispecies specific. Other interfering antibodies may be specific human anti-animal antibodies, which may be produced against animal immunoglobulins (Ig). For example, this process occurs when a patient makes antibody to OKT3, a mouse monoclonal against CD3 used for preventing or reversing allograft transplant rejections.
Heterophile antibodies can interfere by causing false-positive results in two-site immunoassays by bridging the capture and detection antibodies. Likewise, heterophile antibodies can cause false-negative results by binding directly to the capture antibody and thus blocking the reactive site from binding the analyte of interest. Many immunoassays, including ELISA and Luminex-based assays, use animal proteins such as bovine serum albumin and casein to block reactive sites of the microtiter plate or polystyrene microspheres. These blockers provide another potential source of assay interference, as elevated values, false-positive results, and high background readings may occur as a result of heterophile and human anti-animal antibodies binding directly to the blocking protein.
The Luminex Multi-Analyte Profiling (LabMAP) system is a flow cytometry-based instrument that allows multiple analytes to be assayed simultaneously in a single sample (6, 10). The technology uses 5.6-µm-diameter polystyrene particles called microspheres that are internally labeled with two fluorescent dyes. As the microsphere passes through the flow cell, it is interrogated by two lasers. One laser identifies the microsphere on the basis of the ratio of the two fluorophores contained within the microsphere, while the other laser quantitates the amount of analyte bound to the microsphere on the basis of the intensity of the reporter fluorescence. The surface of each microsphere contains multiple carboxyl groups that function as sites for covalent ligand attachment.
In this report, we describe the use of the Luminex system for the development of an immunoassay for the simultaneous quantification of 10 cytokines in human serum. In addition, we have examined 10 heterophile serum samples shown to initially cause falsely elevated cytokine levels in our multiplexed cytokine assay. By identifying which component of the assay the heterophile antibodies were binding to, assay-specific blockers and absorbents were developed to significantly reduce falsely elevated cytokine values.
|
|
|---|
Also included in this study were 25 blood bank samples used as normal controls. All patient samples included in this study were de-identified according to the University of Utah Institutional Review Board approved protocol (protocol no. 7275) to meet the Health Information Portability and Accountability Act patient confidentiality guidelines.
Development of the multiplexed cytokine assay.
The multiplexed cytokine assay was developed by using a standard sandwich capture format which has been described previously (10, 13). Briefly, monoclonal antibodies to human IL-2, IL-4, IL-6, IL-8, IL-10, and TNF-
were purchased from Pharmingen/BD Biosciences (San Diego, Calif.), IL-2r and IL-12 p70 were purchased from R&D Systems (Minneapolis, Minn.), and IL-1ß and IFN-
were purchased from Pierce-Endogen (Rockford, Ill.). These products were used as capture antibodies and coupled to carboxylated Luminex microspheres by using a two-step carbodiimide reaction (15). Internal controls were created by coupling of pooled normal mouse and rat IgG (50 µg/ml; Sigma, St. Louis, Mo.) and pooled mouse serum (100 µg/ml; Sigma) to additional microspheres. A standard curve for each cytokine was made by mixing known concentrations of recombinant human cytokines IL-4, IL-6, IL-8, IL-10, and TNF-
(Pharmingen/BD Biosciences), human cytokines IL-2 and IL-12 cytokine receptor IL-2r (R&D Systems), and human cytokines IFN-
and IL-1ß (Pierce-Endogen) in serum diluent. The patient serum sample was diluted 1:2 (75 µl of serum diluent plus 75 µl of patient serum) and incubated for 10 min on a rotator before the addition of the coupled microspheres to allow the absorption of heterophile antibodies. The cytokine standards, controls, serum samples, and microspheres were incubated for 2 h at room temperature on a rotator by using 96-well filter bottom microtiter plate (Millipore, Bedford, Mass.) to allow subsequent washing. This step was followed by the addition of 100 µl of biotinylated monoclonal antibodies to human IL-2, IL-4, IL-6, IL-8, IL-10, TNF, and IL-2r (Pharmingen/BD Biosciences), IL-12 (R&D Systems), and IL-1ß and IFN-
(Pierce-Endogen) to each well of the microtiter plate. The final concentrations of the biotinylated detection antibodies ranged from 1 to 2.5 µg/ml. For the studies conducted to determine the amount of heterophile binding directly to the specific cytokine monoclonal antibody-coupled microspheres, as well as the mouse and rat protein-coupled internal controls, an anti-human IgG-conjugated R-phycoerythrin detection antibody was used instead. Following a second 30-min incubation on the rotator and washing, 100 µl of streptavidin (10 µg/ml)-conjugated R-phycoerythrin (Molecular Probes, Eugene, Oreg.) was added to each well. After a 15-min incubation and a final wash, the microspheres were resuspended in 100 µl of phosphate-buffered saline with Tween 20 (PBST), and the 96-well microplate was placed in a Luminex 100 instrument with an XY platform (automated microtiter plate handler). The amount of cytokine bound to the microspheres by this antibody sandwich technique is determined by the median fluorescence intensity (MFI) of the reporter molecule, phycoerythrin. The MFI of the unknown sample is then converted into a picograms-per-milliliter value based on the known cytokine concentrations of the standard curve by using a five-parameter regression formula. Since the analyte specificity and position of each microsphere classification in the array is known, a single fluorescent reporter molecule can be used to measure all 10 cytokine concentrations.
Heterophile-blocking serum diluent. Reagents used for the formulation of the serum diluent used in our multiplexed assay for blocking heterophile interference were purchased from Sigma and included fetal bovine serum (FBS; product no. F2442, lot nos. 80K8400, 41K8406, and 52K8411), pooled normal mouse serum (product no. M5905, lot nos. 89H9020 and 073K9305), reagent-grade purified rat IgG (product no. R9759), pooled normal rat serum (product no. I4131), and reagent-grade purified mouse IgG (product no. I5381).
|
|
|---|
![]() View larger version (25K): [in a new window] |
FIG. 1. Mean MFI values for 25 blood bank serum samples (gray bars) and 10 heterophile serum samples (black bars) with mouse and rat protein-coupled microspheres. Reactivities of heterophile samples to mouse IgG, mouse serum, and rat IgG were significant (with P values of 0.0007, 0.001, and 0.0004, respectively) compared with those of blood bank samples. Error bars indicate standard deviations.
|
monoclonal antibody-coupled microsphere, for instance, shows considerable heterophile antibody binding in the PBST diluent (MFI, 1,195), and this binding was greatly reduced by the five blocking diluents, 10% mouse serum (MFI, 91), 10% rat serum (MFI, 170), 10% fetal bovine serum (MFI, 101), nonimmune rat IgG (MFI, 254), and nonimmune mouse IgG (MFI, 149). The most effective blockers appear to be the mouse serum and the nonimmune mouse IgG, which is logical because the IFN-
capture antibody coupled to the bead is a mouse-derived immunoglobulin. This correlation was not always found to be the case, however. For the rat-derived IL-10 monoclonal antibody-coupled microsphere (Table 1), the 10% rat serum diluent had little effect in reducing heterophile binding relative to that occurring with PBST (an MFI of 2,098 versus an MFI of 2,496). The nonimmune rat IgG diluent also seemed to have little effect in reducing binding, with an MFI of 2,014 versus an MFI of 2,469 for PBST. For this specific sample (heterophile sample 4), the most effective blocking diluent was 10% FBS, which reduced the amount of heterophile binding by 86% (an MFI of 2,498 for PBST versus an MFI of 349 for 10% FBS). The specific results for the five blocking diluents for the remaining eight monoclonal antibody-coupled microspheres and three internal controls are shown in Table 1. The results of the blocking diluent study for the rest of the heterophile serum samples are summarized in Table 2. There was an insufficient sample volume to include heterophile sample 8 in this portion of the study. Shown in Table 2 is the most effective blocking diluent for every combination of heterophile sample and coupled microsphere and the percentage of binding reduction. In 48% of the reactions (50 out of 104), the 10% FBS diluent was the most effective diluent in reducing heterophile binding to the coupled microspheres. The serum diluent composed of 10% mouse serum was the second most effective blocker, showing the greatest reduction of heterophile antibody binding in 39 of 104 (37.5%) of the reactions. The 10% rat serum diluent was the third most effective blocker, showing the greatest binding reduction in 15 of 104 (14.4%) of the reactions. The nonimmune rat and mouse IgG-containing diluents were the least effective diluents in reducing heterophile antibody binding. These data show that for a multiplexed assay involving different species of monoclonal capture antibodies, a serum diluent composed of only one species of blocking reagent is not sufficient for reducing heterophile antibody interference. For this reason, titration studies of diluents containing the three most effective blockers, FBS, mouse serum, and rat serum, were conducted. For the cytokine-specific monoclonal antibodies used in our multiplexed assay, the most effective serum diluent for blocking heterophile interference was composed of 10% (vol/vol) FBS, 5% (vol/vol) pooled normal mouse serum, and 2.5% (vol/vol) pooled normal rat serum. Results show that the optimized diluent containing the FBS, mouse serum, and rat serum significantly (P < 0.05) reduced heterophile interference with 8 of the 10 measured cytokines and all three internal controls compared to that occurring with the PBST diluent (Fig. 2). To insure that the blocking effects of the commercial serum were not dependent on a certain production lot, three different lots of FBS and two different lots of mouse and rat sera were examined. Differences in results for the heterophile samples and the blood bank controls between different serum lots were all less than 10% of the coefficient of variation and therefore within normal interassay variance. |
View this table: [in a new window] |
TABLE 1. Comparison of MFI values for five formulations of blocking reagents (three different animal sera and two different nonimmune globulins) to those for PBST to evaluate their effectiveness in reducing heterophile antibody binding to microsphere-bound capture antibodies and proteins for heterophile sample 4
|
|
View this table: [in a new window] |
TABLE 2. Summary of results of heterophile antibody-blocking diluentsa
|
![]() View larger version (31K): [in a new window] |
FIG. 2. Comparison of the mean concentrations or MFI values of 10 heterophile antibody-containing serum samples in a multiplexed cytokine immunoassay. Most reductions in falsely elevated cytokine levels in the optimized serum diluent (gray bars) compared with the levels in PBST (black bars) were significant (IFN- , P = 0.006; IL-10, P = 0.035; IL-12, P = 0.024; TNF- , P = 0.040; IL-1ß, P = 0.062 [not significant]; IL-2, P = 0.006; IL-2r, P = 0.064 [not significant]; IL-4, P = 0.021; IL-6, P = 0.023; IL-8, P = 0.035; mouse IgG, P = 0.013; mouse serum, P = 0.026; rat IgG, P = 0.006). Error bars indicate standard deviations.
|
in the sample diluent versus 315 pg/ml in the PBST. The heterophile antibodies in these samples may bind directly to the reactive site of the IL-2r and TNF-
capture antibodies, thus partially blocking the binding of the cytokine and causing falsely depressed values. While the optimized serum diluent proved to be effective in reducing, and in some cases eliminating, heterophile antibody interference, we also wanted to ensure that the diluent did not adversely affect specific cytokine results by increasing or reducing measurements. To achieve this objective, seven standards containing various concentrations of recombinant cytokines and three controls known to contain native cytokines were run in both diluents. Linear regression and paired type 2 t test analysis for the 10 samples (serum diluent versus PBST) were carried out. The R2 values were all 0.994 and greater (average, 0.999), and the slopes ranged from 0.96 to 1.45 (average, 1.096). The P values ranged from 0.110 to 0.684, with an average of 0.333. These data show that there is no significant difference between results achieved with the two diluents, indicating that the serum diluent does not adversely affect the measurement of known quantities of cytokines.
|
|
|---|
While the described serum diluent proved to be very effective in reducing heterophile interference in our limited sample population, a single formulation may not be effective for all individuals because the effects of the sample matrix are often dependent on the particular individuals involved and their various exposures to different heterophile-inducing immunogens. It has in fact been demonstrated that not all heterophile interference can be blocked by nonimmune globulin, even if pooled globulins from several species are used (5, 10, 16). It is for this reason that we included the rat and mouse IgG and mouse serum microsphere-coupled internal controls in our multiplexed serum cytokine assay. These controls allow the detection of potential interfering antibodies in each individual patient sample that are not eliminated by the normal animal serums contained in our optimal serum diluent.
The multiplexing capability of the Luminex instrument is proving to be a powerful platform for the development of multiple analyte profiles which require lower patient sample volumes and fewer reagents and cost less than traditional methodologies. We have found the Luminex instrument to be an accurate and reliable system for simultaneously quantitating 10 cytokines in only 75 µl of patient serum. Determining exactly which components of the assay were being affected by heterophile interference allowed the addition of true internal controls to each individual patient assay. Because of the efficiency and lower costs of the multiplexed assay, cytokine profile testing is now more feasible to perform in a clinical laboratory setting. Studies are increasingly showing correlations between cytokine production and immunologic and inflammatory disorders in numerous disease processes. With the ability to measure multiple cytokines from a single sample source, we now have a powerful research and clinical tool for assessing cytokine profiles in patient serum samples.
|
|
|---|
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||